FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
31
of
146
USB Port
The micro USB connector is located on the rear of the instrument. The usage of the USB 2.0 and its
specifications are shown in the Table 4-11. Refer to
Usage
Interfacing with PC via USB 2.0
Data formats
Continuous, Fast Continuous, BSI Protocol to PC
near the instrument
Table 4-11
– Data formats of the USB port
Installation of Alibi SD card
Alibi memory requires the Alibi SD card at the SD1 card slot on the bottom of the main board as shown in the pictures
below. If you install the alibi memory SD card, order Alibi SD pack from Flintec or dealer.
Figure 4.28- Location of Alibi SD card slot
Inserting or removing the Alibi SD card
1. Deenergize the instrument and wait 30 seconds before interfering the instrument.
2. Remove 4 pcs M4 screws at corners of the backplane.
3. Draw out the back plane and mainboard from the body of the enclosure. The location of the alibi
card is on the rear side of the main board which is named as SD2.
4. Insert the Alibi SD card into the slot until it locks into place. Or remove the SD card by pressing it
gently. Do not use screw driver to install or remove SD card.
5. Place the sealing sticker as indicated at the sealing section.
6. Place the main card in to the instrument and mount the backplane to the body.
7. Energize the instrument.