The installation of the power supply is carried out by the operator. The operator has to provide an overcurrent protector
(fuse or similar device) disconnecting all energizing wires in case of an inadmissible high current consumption. The
impedance of the protective earth has to be low ohmic in order not to allow touch voltage pass the permissible limit. The
equipotential bonding terminal serves as functional earth of the transmitter.
• Connect the power cable to the transmitter (see section 7.2.1, Fig. 7.9 and Tab. 7.6).
Important!
The degree of protection of the transmitter will only be guaranteed if the power cable fits firmly and tightly in the cable
gland.