7 Connection
72
UMFLUXUS_F7V4-6-2EN, 2017-10-01
FLUXUS F705, F706
• Remove the second blind plug on the right for the connection of the extension cable (see Fig. 7.27).
• Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.28 and Tab. 7.12).
• Push the extension cable through the cap nut, the compression part, the basic part, the reducer and the gasket ring.
(Sealing ring: only for cable gland M20, not for cable gland 1/2 NPS.)
• Prepare the extension cable.
• Insert the extension cable into the housing.
• Tightly screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Fix the extension cable by tightening the cable gland with the counter nut with the ferrite nut.
• Connect the connectors of the extension cable and the temperature probe.
• Connect the extension cable to the terminals of the transmitter (see Fig. 7.2, Fig. 7.27 and Tab. 7.15)
FLUXUS F709
• Prepare the extension cable (see Fig. 7.31).
• Connect the connectors of the extension cable and the temperature probe.
• Connect the extension cable to the terminals of the transmitter (see Fig. 7.4, Fig. 7.30 and Tab. 7.15).
7.2.3
Connection with Junction Box
For the connection of the extension cable to the transmitter see section 7.2.3.1.
For the connection of the extension cable and the temperature probe to the junction box see section 7.2.3.2.
7.2.3.1
Connection to the Transmitter
FLUXUS F704
• Remove the second blind plug on the right for the connection of the extension cable (see Fig. 7.25).
• Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.26 and Tab. 7.12).
• Push the extension cable through the cap nut, the compression part, the basic part, and the reducer.
• Prepare the extension cable.
• Insert the extension cable into the housing.
• Screw the gasket ring side of the reducer tightly into the housing of the transmitter.
• Tightly screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter (see Fig. 7.1, Fig. 7.25 and Tab. 7.15).
Tab. 7.15: Terminal assignment of the transmitter
terminal
extension cable
T1a...T4a
red
T1A...T4A
gray
T1b...T4b
blue
T1B...T4B
white
S1...S4
shield
Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).