TROUBLE-SHOOTING
If after checking the aforementioned conditions, the unit still does not nebulise, we rec-
ommend that you contact the dealer or an authorised Flaem service centre.
PROBLEM
CAUSE
SOLUTION
The device
switches on but
does not nebulise
or nebulises little.
• Malfunction of the
Mesh (2G)
• Soak the Nebulising head
(2G) in a solution of water
(50%) and white wine vinegar
(50%), rinse and fit back on as
described in paragraph 6.3
SANITISING
• The
Nebuliser head
(2G)
is broken
• Replace it with a new
Nebuli-
ser head (2G)
• No medicine supply
in the nebulising area
•
Verify that the small hole in the
centre of the
Medication cup
cover (2D)
is not obstructed.
If so clean it with a needle
• The nebulising speed
depends on the type
of medicinal product
used
• The inhaling treatment time
depends on the type of
medicinal product and on the
patient’s inhaling ability
The device
switches off after a
few seconds (blue
light).
• No electrical contact
between the
Remov-
a b l e M e d i c a t i o n
chamber (2)
and the
Main Unit (1)
• Check that the
Removable
Medication chamber (2)
is properly hooked to
Main
Unit (1)
The device
switches
off
• It’s been 5 minutes
since device switch-
ing on
•
Turn on the unit again
The
START AND
STOP Button (1B)
shows a flash-
ing red light and
the device does
not nebulise or
switches off.
• Batteries are used
up or low
• Replace the batteries if they
are
alkaline
.
• Recharge the
Batteries
if they
are rechargeable
Ni-MH
or
Lithium
type.
The device does
not switch on.
• No electric power
supply
• Verify that
Batteries
are insert-
ed with the correct polarity.
• Used up or low
Batteries.
Replace or recharge.