12.3 Upper blade adjustment
1) On the above condition, close the lid and tighten the two upper blade adjustment screws gradually
while pressing firmly on the PUSH portion of the lid. Tighten both sides screw alternately until the
full length of upper blade tip to touches lower blade tip.
The clearance between lid and body is approximately 0.1mm
Refer to the next page illustration.
2) Set the upper blade by tightening the set screw while pressing firmly on the PUSH portion of the lid.
Use the torque wrench and tighten the screw with 1.2Kg
・
cm.
*
Same blade shape as upper blade and lower blade.
*
Both blades should be replaced at the same time.
13 Others
13.1 Mounting the Stand (Optional parts) for the unit
1) Remove the Battery cover.
2) Put the stand to the unit, then tighten the screw.
13.2 Storage for the S944 Battery Pack
If S944 Battery Pack needs to be stored for more than 1 month without being used, please store it
using the following procedure.
1) Remove the S944 Battery Pack from S218R-200 Main body, then move that to proper storage
area.
2) Maintain the storage environment under following conditions, temperature: -20°C to +20°C, and
humidity: 15 to 95 %RH without condensing.
3) Charge the Battery approximately 30 minutes every 6 months.
(Caution: If the Battery is over discharge, it may be unusable.)
Upper blade set screw
(2-M2
×
1.5mm)
Lower blade set screw
(2-M2
×
1.5mm)
Upper blade adjustment screw
Adjust the upper blade while pressing PUSH marked position.
Target clearance
Adjust it in about 0.1 mm
clearance (Almost one piece
of paper).
Remove the Battery cover
Screw the stand