7.4 Sensor’s Signal levels
The “Signal” value shall be used when further troubleshooting is required.
Signal Levels
Sensor connected directly into the monitor (without a patch cord or Easy Through)
SIGNAL LEVEL (%) MEANING
Signal > 88
EXCELLENT
80 < Signal < 88
GOOD
Signal < 80
CLEANING or INVESTIGATION IS RECOMMENDED
Sensor with a patch cord and/or Easy Through
SIGNAL LEVEL (%) MEANING
Signal > 80
EXCELLENT
40 < Signal < 80
GOOD
Signal < 40
CLEANING or INVESTIGATION IS RECOMMENDED
7.5 Power Protection Fuse
Should it become necessary to verify or change the 1A protection fuse, proceed as following:
1.
First, remove power from the unit
2.
Using a flat head screw driver or a coin,
unlock the fuse cap by doing a left half-turn
3.
The fuse cap is spring-loaded and will self-
pop-out when totally unscrewed as displayed
4.
You can then verify the state of the fuse and
replace it if necessary
5.
You can then put it back doing the reverse
operation while pushing it back in place
6.
Using a flat head screw driver or a coin, lock
the fuse cap by doing a right half-turn
7.
Put the power back on and validate that it is
now working
The FUSE details
:
1A Fast 250VAC 5X20MM / Littelfuse 0235001.HXP or equivalent