EasyGrid User Guide MAN-00084 R 11.0
60
7.6
Diagnostic
This page displays the FISO system diagnostic values for each channel. It is important to
check the Light and Signal levels during the installation of the system.
Information provided
Left column icons provide Sensor Channel Number, Channel Mode and Channel State
Channel 1
Not Enabled
Enabled & Normal
Enabled & in Failure
Channel
mode &
state
Under Temp appears the live measured temperature value
Under Light %, the icons provide Light diagnostic state
Light %
Excellent
Good
Fail
Light %
state
Light %
range
< 40
< = 80
> 80
Under Light %, the Light % value provides the level at which the system eluminates the channel
o
The “Light %” is related to the required power applied to the light source to generate an
accurate temperature measurement from a sensor. A high value of light % could
indicate one of the following problems:
Contaminated or Dirty connection
Over bent cable along the sensor cable routing
Broken Sensor or Extension Cable
Damaged Sensor Tip