5.5.7.
Replacing Electronic Card 43
•
Disconnect all connectors from Card
43
.
•
Remove Card
43
from the four pivots
44
.
•
Mount the new Card.
5.6 REQUEST FOR ASSISTANCE
In case you have any type of technical problems
you can directly contact Rimec
or your local dealer. This
will ensure you get the most rapid and accurate answer to yourproblem.
6. ELECTRICAL SAFETY TESTS
6.1 DESCRIPTION OF TESTS
The test shown below must be performed with suitable equipment and by skilled personnel.
•
Measurement of the enclosure leakage current (of the power supply unit): by powering the
instrument at the nominal operating voltage, increased by 10%, the leakage current must be lower
than 0.1 mA.
6.2 FREQUENCY OF TESTS
This test must be performed every two years.
7. SPARE PARTS
7.1 ORDERING SPARE PARTS
Spare part orders are to be made directly to Rimec or the local dealer.
The codes of the components and the desired quantity must be indicated.
7.2 IDENTIFICATION OF SPARE PARTS