![FISIOTECH FISIOTEK LT Use And Maintenance Manual Download Page 31](http://html1.mh-extra.com/html/fisiotech/fisiotek-lt/fisiotek-lt_use-and-maintenance-manual_2300960031.webp)
31 of 37
Fig. 24
Fig. 25
5.5.3.
Replacing Safety Microswitch 34
•
Perform the operations indicated in paragraph 5.5.1. to remove The Clamp 6 + Protection disk 32
assembly.
•
Loosen the two screws of Cursor
3
and remove the cover.
•
Unsolder the two cables of Microswitch
34
.
•
Replace Microswitch
34
, solder the cables and protect the contacts with heat-shrinking tape.
5.5.4.
Replacing Toothed Belt 45
•
Loosen the four Screws
35
and remove Toothed belt
45
.
•
Mount the new Toothed belt
45
and adjust its tensioning before tightening the four Screws
35
.