6 | Features
Digital Drybath
Features
This chapter introduces the usage, transportation, storage
conditions of the instrument, as well as its basic parameters,
performance and functions.
Normal Working
Conditions
Transportation and
Storage Conditions
Ambient temperature: -20°C~+55°C
Relative humidity:
≤
80%
Basic Parameters
Ambient temperature
10°C ~ 30°C
Relative humidity
≤
80%
Power supply
100-120 V ~ 50/60 Hz or
200-240 V ~ 50/60 Hz
Operation
Up to 2000 m
Pollution Degree
2
Overvoltage category
II Indoor use
Mains supply voltage
fluctuations
±10%
CAUTION:
Before using the instrument, please
make sure the working condition meets the
above requirements. Especially that the power
line is reliably grounded.
Table 2. 100-120 V
NA Catalog Number
88860021
88860022
88860023
Model Name
Digital 1 block
100-120 V
Digital 2 block
100-120 V
Digital 4 block
100-120 V
Temperature Range
Ambient Tempe5°C ~ 130°C (Ambient Temperature 25°C)
Temperature Uniformity
±1.0°C@37°C
Temperature Accuracy
±0.5°C@37°C
Temperature Fluctuation
±0.5°C
Temperature Rise Time
≤ 20 min (rise from 30°C to 130°C)
≤ 25 min (rise from 30°C to
130°C)
Size (L × W × H)
288 × 200 × 100 mm
11.3 × 7.87 × 3.94 in
318 × 200 × 100 mm
12.5 × 7.87 × 3.94 in
450 × 200 × 100 mm
17.7 × 7.87 × 3.94 in
Weight
2.9 kg (6.39 lbs)
3.3 kg (7.28 lbs)
4.7 kg (10.36 lbs)
Fuse Protector (ø5×20)
F 250 V 2.5 A
F 250 V 5 A
F 250 V 8 A