If your device turns off before the 20 minutes are up, or when the device is set down on a table or lightly jarred or tapped, 
re-insert the batteries. Sometimes the batteries will misalign with the metal receptors in the battery compartment and as a
result, the device will turn off when lightly jarred or tapped.
If the green “ON” light does not illuminate when the device is turned ON, first turn the device completely OFF by rotating 
the dial clockwise until you hear a click, then try turning the device ON again. Replace the batteries with fresh ones if you 
still have trouble.
If the yellow indicator lights (1-4) do not illuminate, be sure the sponges are new, clean and thoroughly wet. The hair / skin 
beneath the sponge electrodes must also be sufficiently wet. People with thick hair may have more difficulty wetting the 
hair enough to cause the electrodes to conduct. Some people with small heads will require a tighter headband to make the 
electrodes work properly – adequate pressure is needed against the electrodes by the headband.
To test if the electrodes are working, press the two sponge electrodes together (like an “Oreo cookie”), turn the device ON (green 
light flashing), and, while continuing to squeeze the sponge electrodes together, continue dialing up (counterclockwise)
until you see the yellow lights illuminate.
If the yellow lights illuminate when you test them but not when you have them in use on your head, then the wetness of the
sponges / hair / scalp is insufficient or the tightness of the headband holding the electrodes in place may be insufficient. Make 
sure the headband provides a snug fit. If the headband is not pressing the electrodes firmly enough, adequate conductivity 
may not occur. Press the wet electrodes firmly while the device is on and see if this causes the yellow lights to illuminate. If 
they illuminate, you will need to tighten the Velcro headband.

If your device beeps throughout the treatment session, or fails to beep at the end of the session, replace the batteries.

For troubleshooting support, call us at (800) 692-4380.

TROUBLESHOOTING

Summary of Contents for CIRCADIA Series

Page 1: ...The Device for Stress and Sleep Management User Guide...

Page 2: ...for Use Assembly 4 Device Controls and Use 5 Troubleshooting 6 Cleaning the Device Battery and Sponge Replacement 7 Warnings and Precautions Limited Warranty 8 Extending Your Trial or Returning Your...

Page 3: ...res Circadia Velcro Headband Trial Sponges Circadia has been scientifically proven to help manage stress and sleep Circadia works by comfortably stimulating the brain to produce serotonin and melatoni...

Page 4: ...e headband on the head so that it runs just above the eyebrows Then place each electrode sponges MUST be very moist beneath the headband just above the sideburns see the following page The bottom of t...

Page 5: ...uminate replace or adjust the batteries Once the green ON light illuminates continue the counterclockwise rotation until the second Level 2 indicator light flashes yellow You may not experience any se...

Page 6: ...ter headband to make the electrodes work properly adequate pressure is needed against the electrodes by the headband Totestiftheelectrodesareworking pressthetwospongeelectrodestogether likean Oreocook...

Page 7: ...Patients should avoid using the device near areas of the body where there is embedded shrapnel or metal plates There is no danger in using the device if you have dental fillings Skin irritation may oc...

Page 8: ...andling or which has not been used in accordance with the enclosed instructions or for the stated purposes This warranty is expressly limited solely to the original purchaser and does not extend to an...

Page 9: ...9 Tel 800 692 4380 Fax 800 657 7362 Email info fisherwallace com CONTACT INFORMATION...

Reviews: