background image

7

PIPETTING OF HETEROGENEOUS SAMPLES

(deproteinization in blood glucose determination, for example)

Use steps 1 and 2 of the forward technique to fill the tip with blood.

Wipe the tip carefully with a dry clean tissue.
1.  Immerse the tip into the reagent and depress the push button to the first stop, making sure

  the tip is well below the surface.

2.  Release the push button slowly to the ready position. This will fill the tip. Keep the tip in 

  the solution.

3.  Depress the push button to the first stop and release slowly. Keep repeating this procedure

  until the interior wall of the tip is clear.

4.  Finally, depress the push button all the way to the second stop to completely empty the tip.

CALIBRATION

All Finnpipettes are factory calibrated and adjusted to give the volumes as specified with 

distilled or deionized water. The pipettes are constructed to permit re-adjustment for liquids of 

different temperature and viscosity.

DEVICE REQUIREMENTS AND TEST CONDITIONS

An analytical balance must be used. The scale graduation value of the balance should be 

chosen according to the selected test volume of the pipette:

Volume range 

readable graduation

under 10 µl 

 

0.00 1 mg

10-100 µl 

 

0.01 mg

above 100 µl 

 

0.1 mg

Test liquid: Water, distilled or deionized, “grade 3” water conforming ISO 3696. Tests are done 

in a draft-free room at a constant (±0.5°C) temperature of water, pipette and air between 15°C 

to 30°C. The relative humidity must be above 50%. Especially with volumes under 50 µl the 

air humidity should be as high as possible to reduce the effect of evaporation loss. Special 

accessories, such as the evaporation trap, are recommended.

PROCEDURE TO CHECK CALIBRATION

The pipette is checked with the maximum volume (nominal volume) and with the minimum 

volume. A new tip is first pre-wetted 3-5 times and a series of ten pipettings is done with both 

volumes. A pipette is always adjusted for delivery (Ex) of the selected volume. 

Procedure:

1.  Do 10 pipettings with the minimum volume.

2.  Do 10 pipettings with the maximum volume.

3.  Calculate the accuracy (A) and precision (cv) of both series.

4.  Compare the results to the limits in the Table 1.
If the calculated results are within the selected limits, the adjustment of the pipette is correct.

 

 

7

Summary of Contents for Finnpipette II

Page 1: ...1 INSTRUCTIONS FOR USE Digital Variable Fixed Volume Fisherbrand Finnpipette II 2000 PARK LANE Pittsburgh PA 15275 800 766 7000...

Page 2: ...2 A C B 6 A C B 1 2 3 4 4 A C B 7 A C B 5 1 2 11 9 10 3 1 2 3 4 1 2 3 4 1 2 3 4 5 12...

Page 3: ...3 14 A B C 13 8 CONTENTS PRODUCT DESCRIPTION 4 PIPETTE OPERATION 5 PIPETTING TECHNIQUES 5 CALIBRATION 7 MAINTENANCE 9 TROUBLE SHOOTING 11 PACKAGE 11 SPARE PARTS 12 13...

Page 4: ...0 200 Ext 21377818 5 l 50 l 250 Universal 300 200 Ext 21377819 10 l 100 l 250 Universal 300 200 Ext 21377820 20 l 200 l 250 Universal 300 200 Ext 21377821 100 l 1000 l 1000 1000 Ext 21377822 200 l 100...

Page 5: ...gearing mechanism To release the tip point the pipette at suitable waste receptacle and press the tip ejector with your thumb SAFETY LABEL You can mark the pipette application your initials the calib...

Page 6: ...with the liquid to be dispensed 1 Depress the push button all the way to the second stop 2 Dip the tip under the surface of the liquid in the reservoir to a depth of about 1 cm and slowly release the...

Page 7: ...graduation value of the balance should be chosen according to the selected test volume of the pipette Volume range readable graduation under 10 l 0 00 1 mg 10 100 l 0 01 mg above 100 l 0 1 mg Test liq...

Page 8: ...trap or lid on the vessel is used the correction of evaporation is usually unnecessary Range Volume Accuracy Precision l l s d l cv 0 2 2 l 2 0 080 4 0 040 2 0 0 2 0 080 40 0 020 20 0 0 5 10 l 10 0 12...

Page 9: ...ORT TERM CHECKING The pipette should be checked at the beginning of each day for dust and dirt on the outside surfaces of the pipette Particular attention should be paid to the tip cone No other solve...

Page 10: ...to the instructions DISASSEMBLING 50 1000 l 41 1000 l PIPETTES 1 Press the tip ejector 2 Insert the pliers under the ejector bar to release the tip ejector 3 Remove the tip cone using the maintenance...

Page 11: ...between tip and tip cone Foreign particles between the piston the O ring and the cylinder Insufficient amount of grease on cylinder and O ring O ring damaged Incorrect operation Tip incorrectly attach...

Page 12: ...7 15 15 13 13 14 15 16 13 16 15 14 14 5 17 16 17 10 10 1 2 1 5ml 2 5ml Fix 2 10ml 6 10ml Fix 0 2 2 l 1 2 l Fix 5 50 l 11 40 l Fix 2 20 l 10 100 l 14 15 13 14 16 17 13 14 15 16 20 200 l 41 200 l Fix 10...

Page 13: ...13 13 13 14 14 12 10 10 20 10 12 12 11 17 18 21 20 19 11 22 23 21 19 18 18 19 20 10 10 11 11 11 12 12 11 12 12 a b 10 10 11 11 11 12 12 19 18 b 20 15 16 15...

Page 14: ...1131800 20 10593360 21 1030170 Shelf hanger 2206740 10 10593410 11 10593630 12 10593100 13 10589450 14 1130560 15 1130550 16 1054260 17 1030020 20 200 l 41 200 l Fixed 5 10592520 5 10592540 Fix 50 l...

Page 15: ...0034 1 0035 1 0035 1 0036 23 5 1 0034 1 0035 1 0035 1 0036 1 0036 1 0036 1 0037 24 0 1 0035 1 0036 1 0036 1 0037 1 0037 1 0038 1 0038 24 5 1 0037 1 0037 1 0038 1 0038 1 0039 1 0039 1 0039 25 0 1 0038...

Page 16: ...16 1507910 07 For customer service call 1 800 766 7000 To fax an order use 1 800 926 1166 To order online www fishersci com 2002 Fisher Scientific Litho in U S A BN...

Reviews: