Fisher Scientific 88861021 Instruction Manual Download Page 7

The front panel contains all the controls needed to operate the unit.

TIME

This Chapter covers the control panel and its operation.

1.Start/Stop button: Start or stop the instrument.
2.RUN indicator light: The light is on when the instrument is running and
   off when the instrument is in standby.
3.TIME display window: The window shows cumulative time (in continuous

 

mode) or remaining time (in timer mode). The range of time displayed is 0

 

to 99 hours and 59 minutes. The accuracy is 1 minute. 

4.SPEED display window: The window shows set speed (when the instru

 

-ment is in standby) or current speed (when the instrument is running). 

5.Set Time Buttons: UP/DOWN Arrow buttons are used to increase/

 

decrease the set time of the instrument.

.Set Speed Buttons: UP/DOWN Arrow buttons are used to increase/

 

decrease the set speed of the instrument.

6

1.Connect all the components according to the figures shown on page 1-
of this manual. Use grounded power outlet.
2.Press down the power switch on the back right side of the instrument and
put it to the “|” state and then the instrument is in standby.

3

Section 3

 

 

Operation 

3.1 Operations of the Control Panel

Preparation

Section 3

Operation

 

3

-

1

3

-

2

3.2 Settings

Time Settings

1. Continuous mode

Press the “     ” or “     ” arrow button below the TIME display window. When
the number shown on the display window starts flashing, press “     ” arrow
button to decrease the time to 00:00 and then release the button. The time
setting is finished after the number shown on the display window has
flashed twice.

2. Timer mode

Press the “     ” or “     ” arrow button below the TIME display window. When
the number shown on the display window starts flashing, press “     ” or “     ”
arrow button to increase or decrease the time value. Release the button
when the time shown on the display window reaches the set value. The time
setting is finished after the number shown on the display window has
flashed twice.

Speed Settings

Press the “     ” or “     ” arrow button below the SPEED display window. 
When the number shown on the display window starts flashing, press “     ”
or “     ” arrow button to increase or decrease the speed value. Release the
button when the speed shown on the display window reaches the set value.
The speed setting is finished after the number shown on the display window
has flashed twice.
   

Note:

 press the “     ” or “     ” arrow buton for a longer time to accelerate

              the setting.

Run and Stop

1. Continuous Mode

Press

button and the instrument will start running with the specified

settings and the RUN indicator light will be on. The TIME display window
will show the cumulative time and the SPEED display window will show the
current speed. Press “     ”button again and the instrument will slow down
until it stops. The instrument will then be in standby and the two display
windows will show the set values.

“     ”

2. Timer Mode

Fisher Scientific

Fisher Scientific

Multi Platform Shaker

Multi Platform Shaker

Summary of Contents for 88861021

Page 1: ...Multi Platform Shaker Catalog Number 88861021 88861022 Instruction Manual...

Page 2: ...s and maintenance must be performed by qualified service personnel Material in this manual is for information purposes only Fisher Scientific committed to a continuing program of product development a...

Page 3: ...Cable House Control Panel Foot 4 Platform Sunken Screws M4 8 4 Multi Platform Shaker Multi Platform Shaker Figure 1 Inspect package and contents upon receipt of the instruments If the package is seve...

Page 4: ...Orbit Diameter Speed Accuracy Display Mode Display Accuracy 50 to 450 rpm load weight 3 kg 50 to 300 rpm load weight 3 to 4 5 kg Time Timing Range Requirement Overall Dimensions Platform Dimensions P...

Page 5: ...ely cut off power to the unit by disconnecting the power cord from the unit or unplug from the wall outlet Disconnect unit from the power supply prior to maintenance and service Any spills should be r...

Page 6: ...he instrument or even cause damage if placing loads asymmetrically 2 4 Capacity and Speed Conical Flask Size ml Max Speed rpm With Flask Clamp With Sticky Pad 2 5 2 6 Section 2 Overview Section 2 Over...

Page 7: ...starts flashing press arrow button to decrease the time to 00 00 and then release the button The time setting is finished after the number shown on the display window has flashed twice 2 Timer mode Pr...

Page 8: ...f Lower 10 RPM than the set speed and no more than 70 of the set speed More than 10 RPM than the set speed and no more than 130 of the set speed Lower 10 RPM than the set speed and lower than 70 of th...

Page 9: ...n of Test Tube Rack 1 Place the test tube rack holder at the corresponding position on the aluminum platform and fasten the 4 sunk screws 3 Insert test tubes Note To prevent splashing containers shoul...

Page 10: ...wheels to fasten the flasks and tighten the fixed handle Installation of Universal Platform with Adjustable Bars 1 Fasten Universal Platform with Adjustable Bars onto the platform Note The rotation s...

Page 11: ...Connect the power Switch on Replace power adaptor Adjust the weight and position of load decrease rotation speed Contact Fisher Scientific Contact Fisher Scientific Load moving Tray loose Machine cas...

Page 12: ...0ml flask clamp 1000ml flask clamp 2000ml flask clamp 3000ml flask clamp 5000ml flask clamp 140 140 mm 58 46mm 66 43mm 59 38mm 85 55mm 130 77mm 105 59mm 168 91mm 231 120mm 194 109mm 125 122 5mm 100 92...

Page 13: ...Spare Parts Fisher Scientific Fisher Scientific Multi Platform Shaker Multi Platform Shaker When used in laboratory conditions and according to these operation instructions and maintenance this produ...

Reviews: