Due to handling in shipment or to heat distortion with use, the door cam
may require adjustment. To adjust the door cam, perform the following:
1. Open door and remove screws holding latch cover in place.
2. Locate nuts securing tongue on cam shaft.
3. Loosen but do not remove outside nut.
4. Adjust inside nut, one full turn clockwise draws door 1/16” closer to
cabinet when door is closed.
5. Secure cam tongue in place by tightening outside nut.
6. Repeat Steps 1 thru 5 as necessary to get desired door seal.
7. Secure latch cover in place with two screws.
Isotemp Standard Incubators
6-7
Fisher Scientific
Section 6
Service
Adjusting the Door
Cam