The following table lists the physical dimensions for the
3000
and
3000R:
3000R
3000
Sample Loading Height
14.7” (37.3 cm)
14.7” (37.3 cm)
Cover Closed Height
16.5” (41.9 cm)
16.5” (41.9 cm)
Cover Open Height
41” (104.1 cm)
41” (104.1 cm)
Width
30” (76.2 cm)
23.3” (59.1 cm)
Depth
25.8” (65.4 cm)
25.8” (65.4 cm)
A clearance of 8 cm (3 inches) should be provided on each
side of the unit, to ensure proper ventilation. Place the
centrifuge on a clean, dry surface, to make certain that the
suction feet at the bottom grip the surface firmly. Keep the
area beneath the unit free of debris and loose materials.
The resting surface must be level, to ensure quiet,
vibration-free operation. A rigid and stable location is
important. An improperly loaded centrifuge may vibrate or
move.
Warning:
International Electrotechnical Commission
standard 1010 part 2-20 limits the permitted movement
of a laboratory centrifuge to 300 mm (12 in) in the unlikely
event of a disruption. Laboratory management procedures
should require that no person or any hazardous materials
enter within this boundary while the centrifuge operates.
MARATHON Series Operation Manual