5 Evaluation of a test series (Statistics function)
Prerequisite:
At least two readings must be saved in the measurement data memory.
Note
: The measurement data memory must always be empty before starting a test series that is to be evaluated
(cf. chapter 7 "Deleting measurements", page 15). If at least two readings are not stored in memory when you first
press
RES
and then [OK], the statistical evaluation will NOT start.
Statistical evaluation of a test series:
1. The display menu shows "
RES
". This is the default setting of the instrument. If
RES
is not currently shown on
the display, use the arrow key [
] or [
] to switch to this setting.
2.
Press button [OK] repeatedly.
After each [OK], the display will change to the next statistic number (see table below).
3. Press [ESC] or do further measurements to continue a block. This allows to do statistical evaluations within a
test series.