FISCHER Mess- und Regeltechnik GmbH
Technical data | 6
BA_EN_DE49_0
21 / 32
Temperature coefficient (TK)
Measuring range
TK zero-point
[% FS/10K]
TK span
[% FS/10K]
mbar
typ.
max.
typ.
max.
0…4
0.2
1.0
0.3
1.0
0…6
0.2
1.0
0.3
1.0
0…10
0.2
0.4
0.3
0.3
0…16
0.2
0.4
0.3
0.3
0…25
0.2
0.4
0.3
0.3
0…40
0.2
0.4
0.3
0.3
0…60
0.2
0.4
0.3
0.3
0…100
0.2
0.4
0.3
0.3
0…160
0.2
0.4
0.3
0.3
0…250
0.2
0.4
0.3
0.3
±2.5
0.2
1.0
0.3
1.0
±4
0.2
0.5
0.3
0.5
±6
0.2
0.4
0.3
0.3
±10
0.2
0.4
0.3
0.3
±16
0.2
0.4
0.3
0.3
±25
0.2
0.4
0.3
0.3
±40
0.2
0.4
0.3
0.3
±60
0.2
0.4
0.3
0.3
±100
0.2
0.4
0.3
0.3
With reference to the basic measuring range (FS), Compensation range
0..60°C.
6.5 Display and control elements
Display
4-digit LC display stating the measuring unit
Keyboard
Foil keypad with 3 buttons
Programming
Attenuation
0.0 … 100.0 s (jump response time 10 / 90 %) for sig-
nal output; separately also for display
Measuring range unit
mbar, bar, Pa, KPa, PSI and inWc
Start / end of measuring
range
User-definable within the basic measuring range
(1)
Output signal
can be set anywhere within the signal range
Zero-point stabilising
Zero-point window max. ⅓ of the basic measuring
range
(2)
Zero point correction
±⅓ of the basic measuring range
(3)
Implementation of char-
acteristic curve
Linear, rooted, max. 4:1 spread, inverted
Password
1 ... 999 (0 = no password protection)
(1) Max. effective spread 4:1
(2) measured values around zero are set to zero.
(3) Zero-point correction to compensate different installation positions.