Technical Data | 6
BA_EN_DA01_VUW
19 / 32
6.2 Input variables
Measuring variable
Differential pressure in gaseous and fluid aggressive media.
General
Rated pressure of the measuring
system
Max. static operating pressure
Durability
One-sided over-pressure-proof up to the
rated pressure of the measuring system
resistance to under-pressure on the (+) and
(-) side
Measurement accuracy
±1.6 % of the measuring range
Temperature sensor
0.3 % / 10 °C
Zero-point adjustment
±25 % of the measuring range
Measuring ranges
Small measuring cell
Ø75
Measuring range
Device model
V
U
W
0 … 0.6 bar
●
●
0 … 1 bar
●
●
0 … 1.6 bar
●
●
0 … 2.5bar
●
●
0 … 4.0 bar
●
●
0 … 6 bar
●
●
0 … 10 bar
●
●
0 … 16 bar
●
●
0 … 25 bar
●
●
-1 … 0.6 bar
●
●
-1 … 1.5 bar
●
●
-1 … 3 bar
●
●
-1 … 5 bar
●
●
Large measuring cell
Ø130
Measuring range
Device model
V
U
W
0 … 40 mbar
●
0 … 60 mbar
●
0 … 100 mbar
●
0 … 160 mbar
●
0 … 250 mbar
●
0 … 400 mbar
●
0 … 600 mbar
●
6.3 Operating conditions
Permissible ambient temperature
-20 … +80 °C
Admissible storage temperature
-40 … +80 °C
Admissible media temperature
Max. 100 °C
Type of protection:
IP 65 acc. to EN 60529