FLS M9.08

DUAL-PARAMETER 

pH/ORP AND 

FLOW MONITOR & TRANSMITTER

SAFETY INSTRUCTIONS

PACKING LIST

General Statements

• Do not install and service the product without following the Instruction 

Manual.

• This item is designed to be connected to other instruments which can be 

hazardous if used improperly. Read and follow all associated instrument 

manuals before using with it.

• Product installation and wiring connections should only be performed by 

qualified staff.

• Do not modify product construction.

Installation and Commissioning Statements

 

• Remove power to the instrument before wiring input and output connections.

• Do not exceed maximum specifications using the instrument.

• To clean the unit, use only chemical compatible products.

Please verify that the product is complete and without any damage. 

The following items must be included:

• M9.08 Dual-Parameter pH/ORP and Flow Monitor & Transmitter

• Instruction Manual for M9.08 Dual-Parameter pH/ORP  

and Flow Monitor & Transmitter

Summary of Contents for FLS M9.08

Page 1: ...allation and wiring connections should only be performed by qualified staff Do not modify product construction Installation and Commissioning Statements Remove power to the instrument before wiring input and output connections Do not exceed maximum specifications using the instrument To clean the unit use only chemical compatible products Please verify that the product is complete and without any ...

Page 2: ...alues to a external device A proper combination of digital outputs allows customized setups for any process to be controlled F3 00 F3 20 F6 30 F3 10 F3 05 F6 60 F6 61 F111 M9 08 X X X X X X ULF F3 80 pH ORP200 pH ORP400 pH ORP600 pH ORP800 C150 200 C100 C300 C6 30 M9 08 X X X X X X TECHNICAL DATA General Associated sensors FLS pH ORP sensors FLS hall effect flow sensors or FLS F6 60 flow magmeters...

Page 3: ...H ORP user selectable as ON OFF Proportional frequency output Timed Pulse MIN alarm MAX alarm Off optically isolated 50 mA MAX sink 24 VDC MAX pull up voltage max pulse min 300 hysteresis User selectable 2 x Relay output Flow user selectable as MIN alarm MAX alarm Pulse Out Window alarm Off pH ORP user selectable as ON OFF Proportional frequency output Timed Pulse Off mechanical SPDT contact expec...

Page 4: ...g for panel or wall installation The panel version is installed using the panel mounting kit M9 SN1 while the wall mounting version is got fixing the panel mounting version on the wall mounting kit M9 KW1 The mounting kits can be ordered directly connected to the monitor or separately and then simply installed on it PANEL MOUNTING WALL MOUNTING ...

Page 5: ...ils M9 SN1 Wall installation Use the panel mounting kit M9 SN1 to fix the M9 08 on the dedicated frontal cutout of the wall mounting kit M9 KWX Tighten front screws of box and waterproof connectors of cables internally mount caps on screw sites to get a IP65 watertight installation ...

Page 6: ...upper part of the terminals for an easy cabling Insert wire tip or ferrule completely into the terminal and fix with the screw until finger tight Do not route the sensor DC power or 4 20mA cables in conduit containing AC power wiring Electrical noise may interfere with sensor signal Routing the sensor cable in grounded metal conduit can help prevent electrical noise and mechanical damage Wall Inst...

Page 7: ... LOOP 1 LOOP 2 LOOP 1 21 22 23 24 Power Supply 12 24 VDC 12 24 VDC PLC 4 20 mA Input 2 4 20 mA Input 1 VDC VDC 2 1 LOOP 2 LOOP 1 LOOP 2 LOOP 1 21 22 23 24 Power Supply 12 24 VDC 12 24 VDC PLC 4 20 mA Input 2 4 20 mA Input 1 V 2 1 L L L L 21 22 23 24 r Supply 4 VDC 4 VDC PLC 0 mA ut 2 0 mA ut 1 VDC VDC 2 1 LOOP 2 LOOP 1 LOOP 2 LOOP 1 21 22 23 24 Power Supply 12 24 VDC 12 24 VDC PLC 4 20 mA Input 2 ...

Page 8: ... Digital INPUT Digital INPUT Imax 50mA lmax 50mA lmax 50mA PLC DIGITAL INPUT N DIGITAL INPUT 2 DIGITAL INPUT 1 REF PLC N O Imax 50 mA Imax COM 3 4 N O 3 4 COM AC or DC Power User Imax 50mA N O 3 4 COM AC or DC Power User Imax 50mA lmax 50mA lmax 50mA N O 3 4 COM AC or DC Power User Imax 50mA N O 3 4 COM AC or DC Power User Imax 50 lmax 50mA lmax 50mA Pt100 Pt1000 no connection 27 28 29 30 PT100 10...

Page 9: ...D REED 12 24 VDC OPERATIONAL OVERVIEW The M9 08 Dual Parameter pH ORP and Flow Monitor Transmitter features a full graphic display and a five button keypad for system set up calibration and operation Full graphic display has a white backlight during standard conditions a green backlight in case a external device control is activated ON OFF PROPORTIONAL FREQUENCY PROPORTIONAL PULSE and TIMED PULSE ...

Page 10: ...re sensor connected NTC not temperature compensation ATC automatic temperature compensation VIEW LEVEL MENU DIRECTORY pH ORP temperature Output Settings item code software release pH ORP temperature flow pH ORP flow pH ORP or temperature or flow analog output 1 pH ORP or temperature or flow analog output 2 pH ORP direct access to calibration flow infinte totalizer resettable totalizer pH ORP last ...

Page 11: ...ld Backlight Reed Bi directional Password Contrast Decimal Point Measurement Decimal Point Flow Default Data Outputs 1 SSR 2 SSR 3 RELAY 4 RELAY Output Test 4 20mA1 4 20mA2 Flow Calibration Correction Factor Signal Intensity BLE Auto Calibration Calibration pH ORP Probe Calibration Temperature Probe Calibration Flow Settings K Factor Volume Unit Flow Unit Settings Manual Temperature Probe Unit Tem...

Page 12: ...ilure a reference number correlated to the involved digital output is reported by the output number PROCEDURE FOR OUTPUTS SETTING go to the Options menu enter into the Outputs activation sub menu enable output s go to the Outputs menu set the operating mode for each enabled output EDIT LEVEL PUSH BUTTON to modify an item to scroll right to return to Menu without saving to save new settings Monitor...

Page 13: ...t relaxed Output energized pH ORP Time Set Point I His Set Point I Prop Band Relay Prop pH ORP Set Point I His Set Point I pH ORP Relay Timed Endpoint 100 pulses from 0 to max pulses min Starting point 0 pulses 5 10 Set Point I His Set Point I Prop Band Relay Prop pH ORP Set Point I His pH ORP Set Point I Relay Timed Hysteresis Setpoint Output relaxed Output energized pH ORP Time Digital outputs c...

Page 14: ...00 W In case of combination with F3 00 W LOW BATTERY status and NO SIGNAL condition can be remoted by two different digital outputs or by one for both indications SOFTWARE UPDATING In order to update the Instrument Software with a New Firmware Release follow the suggested procedures TO UPDATE INSTALLED UNITS Download the interface software FLS Calibration System and the Updated Software on www fls...

Page 15: ...ts 1 5 meter long Part No Name Description M9 SN1 Fixing snails 2 fixing snails for panel installation of FLS monitors Connect M9 08 to the laptop by the USB cable Select the item M9 08 which appears on the navigation area on the FLS Calibration System software Confirm FW UPGRADE and select the Updated Software NOTE At the end of the procedure restart the instruments in order to refresh M9 08 soft...

Page 16: ...FIP Formatura Iniezione Polimeri S p A Loc Pian di Parata 16015 Casella Genova Italy Tel 39 010 96211 Fax 39 010 9621209 www flsnet it I2461 IMM908E REV 02 02 2020 ...