Instructions, Maintenance and Service Manual ULTRAFREEZER Series
11
!
SETP.’
and
▲
or
▼
at the same time. Release the buttons when the correct value is
reached. The display returns to its normal status when the
‘PRG’
button is pressed or when
during 30 seconds no button is used.
When again the
▲
button is pressed, the display will show
P82
. By pressing
‘SETP.’
the
display will show the set value of the
‘actual day’.
This can be changed by pressing the
SETP.’
and
▲
or
▼
at the same time. Release the buttons when the correct value is
reached. The display returns to its normal status when the
‘PRG’
button is pressed or when
during 30 seconds no button is used.
When again the
▲
button is pressed, the display will show
P83
. By pressing
‘SETP.’
the
display will show the set value of the
‘actual hour’.
This can be changed by pressing the
SETP.’
and
▲
or
▼
at the same time. Release the buttons when the correct value is
reached. The display returns to its normal status when the
‘PRG’
button is pressed or when
during 30 seconds no button is used.
When again the
▲
button is pressed, the display will show
P84
. By pressing
‘SETP.’
the
display will show the set value of the
‘actual minute’.
This can be changed by pressing the
SETP.’
and
▲
or
▼
at the same time. Release the buttons when the correct value is
reached. The display returns to its normal status when the
‘PRG’
button is pressed or when
during 30 seconds no button is used.
7.7. Alarm test function
Enter the password, standard factory setting 1234.
After activation, press simultaneously
‘BATT.-MUTE’
and
▲
. All leds, buzzer and
potential free contact will be activated.
The display returns to its normal status when the buttons are released and when during
30 seconds no button is used.
7.8. Changing password
The customer can change the password by pressing first the
‘PRG’
button (LED under
Prg
button light on) and second 4 times the
▲
button. The display will show
P90
. A new password
can be entered by using the following buttons:
▲
= 1;
Setp.
= 2;
Batt.
= 3,
▼
= 4;
Prg
= 5 and
Code
= 6. The password can be modified pressing simultaneously ‘
SETP.’
button and the new
password. The display returns to its normal status when the
‘PRG’
button is pressed or when
during 30 seconds no button is used.
Important
Remember to take note of the password! The freezer cannot be switched off or the
parameters can not be changed or visualized on the display without it. When you forgot the
code, Fiocchetti can supply you a master code. With this master code, all parameters
including the access code can be adjusted.
7.9. Door contact
This is only valid for the upright models. When the door is opened, an alarm delay-time will
start. When this alarm time-delay has passed, an alarm will be activated (E6). The alarm time-
delay is set to 30 seconds.