Fillauer Aeris
Performance 2 Product Manual Download Page 3

3

Contents

Intended Use  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

4

Warnings and Precautions    .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

5

Alignment (Specifications & Preparations Before Use)  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

6

Consumable Components: Foot Shell and Spectra

®

 Sock   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

8

Compatibility   .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

9

Disposal / Waste Handling    .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

9

Warranty    .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

9

User Instructions    .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

10

Serious Incidents    .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .

11

To see other language options, visit 

fillauer.com

 .

Summary of Contents for Aeris Performance 2

Page 1: ...Aeris Performance Product Manual ...

Page 2: ......

Page 3: ...s 5 Alignment Specifications Preparations Before Use 6 Consumable Components Foot Shell and Spectra Sock 8 Compatibility 9 Disposal Waste Handling 9 Warranty 9 User Instructions 10 Serious Incidents 11 To see other language options visit fillauer com ...

Page 4: ...Split Toe Design Pylon Main Spring Pylon Top Spring Figure 1 Indications Moderate to active transtibial or transfemoral amputees as defined by functional K3 and K4 levels Unilateral or bilateral patients Patients that would benefit from greater energy return and multi axial function Patients weighing up to 330 lbs 150 kg Contraindications Build height below 6 75 in 17 cm Patients weighing over 330...

Page 5: ...c orthotic component must not be subjected to dust debris liquids other than fresh water abrasives vibration activities which would damage the biological limb or prolonged extreme temperatures 5 C or 50 C Do not allow debris or liquids to remain in the prosthesis and its components during use Rinse the foot with fresh water and dry immediately after exposure CAUTION The foot unit is waterproof to ...

Page 6: ...eceiver Torque all set screws to the setting specified by the manufacturer of the pyramid receiver For Fillauer components this is 15 N m Proper thread locker must be used for final delivery per the component manufacturer s specifications 5 Figure 2 Static and Bench Alignment Standard bench alignment techniques may be used for the Aeris Performance 2 foot Figure 2 Before aligning the initial heel ...

Page 7: ... and optimize gait Patient feedback during this process is essential Adjustments of the plantar and dorsiflexion angles will help the patient achieve a smooth transition from heel to toe The pylon should remain vertical in the frontal plane throughout gait Procedure Check for smoothness of gait and ground contact throughout the stance phase of gait If the tibial progression is slowed from heel str...

Page 8: ... of the foot shell to maintain its appearance and durability Always use the shell with an internal Spectra sock included Never use a sharp edged tool such as a screwdriver to install or remove the foot shell Installation Slide the provided Spectra sock onto the foot from toe to heel pulling excess material to the ankle so that it does not bunch under the heel or toe of the foot Insert the forefoot...

Page 9: ... hand If the foot shell is too tight a smooth edged shoehorn may be used to disengage the heel lock Compatibility Fillauer feet are appropriate for use with Fillauer or equal ISO 10328 compliant endoskeletal components A Fillauer foot shell should be used with this device the fit of other manufacturers shells cannot be guaranteed Disposal Waste Handling The product must be disposed of in accordanc...

Page 10: ...f the foot is submerged the foot and foot shell should be rinsed with fresh water and dried immediately to remove salt chlorine or debris CAUTION The foot shell is designed to provide realistic appearance and maximum performance of the Aeris Performance 2 The life of the foot shell will depend on level of activity and degree to which it is protected from wear and damage with socks and shoes Socks ...

Page 11: ...Serious Incidents In the unlikely event of a failure resulting in a fall and or injury seek immediate medical help and contact your prosthetist at the earliest possible convenience ...

Page 12: ...710 Amnicola Highway Chattanooga TN 37406 423 624 0946 3938 S 300 W Salt Lake City UT 84107 801 281 9964 Fillauer Europe Kung Hans väg 2 192 68 Sollentuna Sweden 46 0 8 505 332 00 EC REP 2021 Fillauer LLC M058 06 29 12 10 04 21 Rev 1 ...

Reviews: