Fike 203-0003 Installation And Maintenance Instructions Download Page 3

26-0350 Issue 11

 

 
Auxiliary Input / Output 

 
Where Input or Output functionality is required other than at the control panel, it is recommended that a separate 
Sita Loop I/O Module is fitted to the addressable loop as a dedicated I/O module. Where this is not possible, a 
limited auxiliary I/O facility is available from the detector via the auxiliary I/O terminals with the addition of a separate 
Sita Multipoint Boxed Auxiliary I/O Module (Fike part no. 803-0005) or Sita Multipoint Base Auxiliary I/O Module 
(Fike part no. 803-0003). Either an input or an output may be used or controlled respectively through the additional 
I/O module which shares the address of the detector to which it is connected. Please see the relevant panel 
programming manual for details of the options available for different modes of operation. Note that the detector 
auxiliary I/O feature can only be used with the Sita Multipoint Boxed or Base Auxiliary I/O Module. No other 
input/output components or equipment may be connected directly to the detector otherwise damage may occur. 
Note also that sound patterns SP4 and SP6 may not be used if the detector I/O facility is used. 
 

Remote LED 

 
The remote LED terminals may be used to connect a separate external LED unit (Fike 
part  no.  600-0092)  to  the  detector  that  will  activate  only  when  the  device  to  which  it  is 
connected enters an alarm condition. The Fike LED unit incorporates the necessary 10k 
Ohm current limiting resistor. No other components or equipment may be connected to 
the detector remote LED terminals. Connection of other components or equipment may 
damage the detector. 

 

Head Contamination Warning 

 
Warning of head contamination is in two stages. Firstly in the form of a ‘Quiescent optical level high or low’ event 
recorded into the event log and fault or warnings list (depending on the control panel being used), and secondly in 
the form of a ‘Smoke sensor failed; signal high event, recorded into the event log and fault

 

list. 

 
If  the  smoke  detection  mode  is  set  to  SM0  (off)  the  head  contamination may  still  be  monitored  at  the  detector 
depending on the smoke mode, but not by the panel. If contamination is present then the result will be an LED flash 
every five seconds at the detector, but the panel will show no fault. Thus, if a device is being re-programmed to re-
enable a smoke detection mode which has been turned off, check the LED status of the device first and ensure that 
a clean optical chamber is present. If this is not checked and the smoke detection mode is re-enabled, the device 
could signal a fire alarm when re-configured. 
 

Replacing Contaminated Optical Chambers 

 
When  an  optical  chamber  becomes  contaminated  we  recommend  replacing  it  with  a  new  replacement  item. 
Cleaning an old chamber will not give satisfactory results and the reassembly may very easily result in misalignment 
of the optical path.  
 
Simply remove the old optical chamber with the head removal tool and replace with a new unit. Once the unit has 
been replaced the detector will automatically re-calibrate the optical level for the new chamber 
 

Replacing Multipoint Electronics 

 
If an electronics module requires replacement after the system has been programmed and tested then proceed as 
follows: 
 
Stop  the  loop  at  the  control  panel  before  replacing  the  module  with  a  new  one,  noting  its  serial  number  onto  the 
configuration sheets. Initialise the loop from the control panel and re-configure the loop devices using the 

Re-config

 

command.  This  will  send  the  programmed  settings  from  the  control  panel  to  the  loop  devices.  The  panel  should 
then be reset and the device tested for all programmed modes of operation. 

 

 

Testing 

 
Due to the fact that the smoke detection modes are all ‘thermally enhanced’, it may be noticeable that on test the 
devices  respond  more  slowly  than  may  be  expected.  This  is  perfectly  normal  and  is  due  to the  fact  that  during  a 
smoke test there is no additional heat present to cause the sensitivity to be enhanced.  
 
Please note some makes of heat detector test equipment will not operate with these devices due to the position of 
the thermistor. 

Reviews: