5.
Service and maintenance
The installed H40 telemetry station requires occasional checking of the cable connection
status of the measuring probe and GSM antenna. Especially when installed in the field,
the cabling is stressed by weather and sunlight. Mechanical damage to the cables by
various rodents often occurs. Suitable cable protectors and a well-designed station in-
stallation can ensure trouble-free operation of the station throughout its lifetime.
5.1.
Firmware updates
The manufacturer of the Hydro
Logger H40 maintains updated
firmware versions for most of its
products in the "Firmware and
Utilities" partner zone on its
www.fiedler. company
server.
The partner zone is accessible to
authorized users by logging in to
the "Support" menu.
Firmware updates are performed
from the menu of the MOST pro-
gram, which must be licensed for
service companies. A detailed
description of how to update the
firmware can be found on this
website along with the individual
firmware packages.
Follow these steps to update the firmware:
▪
Connect to the unit whose software you want to update with MOST.
▪
Read the current parameters from the connected telemetry station and save them
as a parameter file as a backup of the current instrument settings. Similarly, back
up archived measurement data.
▪
In MOST, select the "
Firmware
" option in the "
Manufacturing
" menu.
▪
Select the desired data file containing the firmware (*.hex) and load it into the PC
by pressing the "Open" button.
▪
After the new firmware transfer is complete, the unit will automatically restart and
MOST will display the message "Firmware update complete".