20
Instrument displays the preset tare value entered.
Net indicator
appears to indicate the value being displays is net weight.
To cancel preset tare effect: -
Remove all loads from platform then press
[Tare],
or
Enter a zero preset tare value then press
[Set]
.
Notes: -
The pre-determined tare weight entered will be rounded to the nearest division of the instrument.
This does not affect the accuracy of the subsequent weighing and operation.
Manual tare is possible when Repetitive Tare (F13) is set to ON.
6.5 Select the Preferred Function Mode
18
This instrument supports the below function modes. Abbreviation of each function mode is bracketed.
Weighing (
Weigh
).
Piece Count (
Count
).
Action-Tare-Memory
(AtM)
.
Peak Hold Function (
Peak
).
Animal Weighing (
Ani
).
Quick access to Customer & Product Code Setting (
PCd
).
Press
[Func]
until the abbreviation of the desired function mode appears then press
[Print/M+]
to enter.
The working mode employed before powering off will be employed again automatically when re-powered on.
6.6 Weight Units
19
This instrument supports kg, g and lb. To shift among various weight units, press
[Unit]
to shift among various
weight units.
6.6.1 Weight unit gram (g)
Disregarding to the setting of
F9
, weight unit gram (g) is available only when 3 or 4 decimal places (0.000 or
0.0000) is selected in F81. Contact your dealer for more information about this.
6.6.2 Select the preferred weight unit
20
The desired weight units should enable in F9. Press
[Unit]
until the preferred weight unit appears.
6.7 Memory Accumulation
21
There are 2 types of memory accumulation: -
a.
Automatic accumulation mode, and
b.
Manual Accumulation mode.
Maximum accumulation limit is = 8 digits (e.g. 99999999) plus decimal (if any).
Err 28
appears when
maximum accumulation limit is exceeded.
18
Depends on F11 setting.
19
Depends on F9 setting.
20
Changing weight unit during operation will clear all accumulate weight data from memory.
21
Memory accumulation does not support Peak Mode.
Summary of Contents for 19eA