41
9. Action-Tare-Memory (ATM)
9.1 Description of ATM Mode
It means action, then tare, then memory: -
Action = load or remove weight from weighing platform.
Tare = the above weight added on or removed from will be tare
off automatically.
Memory = the above weight will be added to or deducted (in case
of removal) from accumulated memory.
9.2 Basic ATM Settings
1.
Refer to
6.6
on how to select the desired weight unit.
2.
Enter ATM mode.
3.
Press
[Set]
to select Auto Accumulation target then press
[Print/M+]
.
Gross = Gross weight will be accumulated.
Net = Net weight will be accumulated.
4.
Instrument prompts for delay time (second). Delay time is the time
interval (00 ~ 99 second): -
Between a valid stable weight result is obtained and before it
is tare off and accumulated to memory.
Display time of total accumulated weight result (after all
loadings are removed) and before it is clear from print out
memory.
5.
Enter delay time through numeric keys then press
[Print/M+]
to enter.
6.
At this point, ATM mode is ready for use.
9.3 Start using ATM
31
1.
Apply container on platform. Instrument will tare off the weight of the
container.
2.
Apply or removed load on or from platform. The weight result is
displayed for the time interval set forth by above point 2. Then
instrument will accumulate the weight result in memory then clear it
from the display.
Notes: -
Both positive and negative weight will be accumulated to
31
Weight changed less than 10d will not be processed.