![FIBER FOX Mini 6S+ User Manual Download Page 41](http://html1.mh-extra.com/html/fiber-fox/mini-6s/mini-6s_user-manual_2286591041.webp)
39
Symptom
Name
Cause
Remedy
Fiber core
axial offset
There’s dust in
V-grooves and
fiber hammer
Clean the V-grooves and
fiber hammer
Fiber core
angle error
There’s dust in
V-grooves and
fiber hammer
Clean the V-grooves and
fiber hammer
Bad fiber end-
face quality
Check the cleaver
Fiber core
bending
Bad fiber end-
face quality
Check the cleaver
Pre-fuse power
too low or pre-
fuse time too
short.
Increase [Pre-fuse
Power] and/or[Pre-fuse
Time].
The mode
field
diameters
mismatch
Arc power not
adequate
Increase [Pre-fuse
Power] and/or[Pre-fuse
Time].
Dust
combustion
Bad fiber end-face
quality
Check the cleaver
Dust still present
after cleaning fiber
or cleaning arc.
Clean fiber thoroughly or
increase[Cleaning Arc
Time]
Bubbles
Bad fiber end-face
quality
Check the cleaver
Pre-fuse power
too low or pre-fuse
time too short.
Increase [Pre-fuse Power]
and/or[Pre-fuse Time].
Separation
Fiber stuffing too
small
Perform [Arc Calibration].
Pre-fuse power too
high or pre-fuse
time too long.
Decrease [Pre-fuse
Power] and/or [Pre-fuse
Time].
High Splice loss: Cause and solutions
Appendix A