39
Error Message
Cause
Solution
Left Fiber Place
Error
The fiber end-face is placed
on the electrode centerline.
Press RESET button, and reload the fibers
to make sure that fiber end face is placed
between V-groove and the centerline of
electrodes.
Right Fiber Place
Error
Press Motor Distance
Over Limit
The fiber is not set correctly
at the bottom of the
V-groove, which results in
that the fiber offsets beyond
motor formation range.
Press the “RESET” button; correctly re-
position the fiber at the bottom of the
V-groove.
Press Motor Error
Motor might be damaged.
Consult your nearest sales agent.
Search Fiber End
face Failed
The fiber is not set
correctly at the bottom of
the V-groove.
Press the “RESET” button; correctly re-
position the fiber at the bottom of the
V-groove.
Arc Failure
Arc discharge does not
occur.
Confirm the electrodes are in proper
position; Replace the electrodes.
Align Motor Distance
Over Limit
The fiber is not set correctly
at the bottom of the
V-groove.
Press the “RESET” button; correctly re-
position the fiber at the bottom of the
V-groove.
Search Fiber Clad
Failed
The fiber is not set
correctly at the bottom of
the V-groove.
Press the “RESET” button; correctly re-
position the fiber at the bottom of the
V-groove.
Fiber Border
Position Wrong
There’s dust or dirt on the
fiber surface
Prepare the fiber (stripping, cleaning, and
cleaving) again; clean the lens; redo [Dust
Check]
Unknown Fiber Type
There’s dust or dirt on the
fiber surface
Prepare the fiber (stripping,
cleaning, and cleaving) again.
Different fiber types on two
sides.
Use an appropriate splice mode other than
AUTO splice mode to re -splice.
Non-standard optical fibers AUTO splice mode can only detect the
standard SM, MM, NZ fibers.
Fiber Clad Over
Limit
Fiber edge is not located in
the camera range.
Adjust the fiber position and perform “motor
calibration” .
Appendix B
During the splice operating process, if the error messages are shown on the screen,
please follow the solution precisely as shown in the list below. If it is not possible to
solve the problem, the splicer may require service by a qualified service center. In
this case, please contact sales agents.
High Splice loss: Cause and solutions