Display for magnetic band
F7
Data: 13/01/16
F7_ing.doc
pag 4/12
This parameter together with the following allows the programming of the value on the display for
a certain displacement of the sensor. It means that is necessary to set the value which has to be
displayed corresponding to a displacement of the sensor on the magnetic band of 10mm.
The Factory value is
VISUAL
=10.0, which is the necessary value to read the displacement in
millimetres with decimal resolution.
The range allowed is from 0,0001 to 99999 with setting of decimal point position that is, after
programming of the last digit on the left, pressing key
will blink the decimal pinpoint and with
key it can be moved to the wanted position. Confirm with
RESET
.
+%
+%
+%
+%
Number of decimal digits
It is the number of decimal digits to visualize on the display, range allowed from 0 to 4.
Example: for every 10mm of linear displacement it is necessary to display 12.345 with only two
decimal places: Set
0
0
0
0
=12,345 and
+%
+%
+%
+%
=2
//
//
//
//
Keys opening mode
This parameter programmes the functions linked to the keys.
The value to set is a number of three digits so each key corresponds to a digit; the digit on the
right stands for setting of key
RESET
, the digit in the middle stands for key while the last digit on
the left stand for key
.
The values allowed are the following:
VALUE
ENTER
/
RESET
0
Not active
Not active
Not active
1
mm/inch conversion
abs/rel
measurement
Reset
2
Not active
Not active
Preset
3
Not active
Not active
Fast Preset
4
Not active
Not active
Change of origin
0,1,2
5
Not active
Not active
Offset
6
Not active
Not active
Fast Offset
7
Not active
Not active
Delayed reset *
(after 3s)
8
Not active
Not active
Delayed preset *
(after 3s)
* Delayed means that to activate the function (reset or preset), it is necessary to keep
ENTER
/
RESET
pressed for approximately 3 s.
Reset:
reset function
of the displayed value; after pressing
ENTER
/
RESET
,
the displayed value is
set to zero.
Preset:
preset function of the displayed value; after pressing
ENTER
/
RESET
,
the displayed value
becomes equal to the value in the
3U6W
3U6W
3U6W
3U6W
parameter, which can be set immediately after
W$6W,
W$6W,
W$6W,
W$6W,
(when the digit on the right is set to 2).
Fast
Preset:
the fast setting of the displayed value on the display; after pressing
ENTER
/
RESET
3U6W
3U6W
3U6W
3U6W
appears and pressing 2 times
ENTER
/
RESET
it is possible to set the value directly (use and
and then confirm with
ENTER
/
RESET
). This function is useful when the displayed value must be
corrected often.
Origin change
(correction of the displayed value for tool changes): with this function 3 different
origins (0,1,2) can be set and with
ENTER
/
RESET
it is possible to switch from one origin to another.
After setting 4 in the first digit on the right of the
W$6W,
W$6W,
W$6W,
W$6W,
parameter,
3U6
3U6
3U6
3U6
will appear and after
Summary of Contents for F7-
Page 8: ...Display for magnetic band F7 Data 13 01 16 F7_ing doc pag 8 12 F7 E AM20 ...
Page 9: ...Display for magnetic band F7 Data 13 01 16 F7_ing doc pag 9 12 F7 I AM20 ...
Page 10: ...Display for magnetic band F7 Data 13 01 16 F7_ing doc pag 10 12 ...
Page 11: ...Display for magnetic band F7 Data 13 01 16 F7_ing doc pag 11 12 ...