Display for magnetic band
F20
Data: 13/01/16
file: F20_ing.doc
pag 5/8
%.+
%.+
%.+
%.+
Count direction
Set out the count direction of the display, range allowed 0 or 1. Factory default 0.
Setting 0, the value on the display increases moving the sensor to the right (sensor as per draw on page 1).
Setting 1, the value on the display increases moving the sensor to the left.
//
//
//
//
Keys opening mode
This parameter programmes the functions linked to the keys.
The value to set is a number of three digits so each key corresponds to a digit; the digit on the right stands for
setting of key
RESET
, the digit in the middle stands for key while the last digit on the left stand for key
The Factory default is 101 or rather with the functions of absolute/relative quote (
ABS
/
REL
key)
and reset
(
RESET
/
ENTER
key)
activated.
The values allowed are the following:
VALUE
KEY
KEY
KEY
RESET
/
ENTER
0
Not open
Not open
Not open
1
Function
ABS
/
REL
Conversion mm/inch
Reset
2
Not open
Display in degrees
Preset
3
Not open
Not open
Fast Preset
4
Not open
Not open
Change of origin1,2,3,4,5
Reset:
function
of reset of quota, pressing on key
RESET
/
ENTER
the quota is zero-set.
Preset:
function of preset of quota, pressing on key
RESET
/
ENTER
the quota on the display became the same
of the one set in parameter Preset. The setting of Preset value appears immediately after parameter
//
//
//
//
(if chose value 2).
Preset Veloce:
the fast setting of the quota on the display, pressing on key
RESET
/
ENTER
appears Preset and
pressing still 2 times
RESET
/
ENTER
is possible to set the value directly (use keys
and confirm with
RESET
/
ENTER
). This function is useful when the quote on the display has often be corrected.
Origin change
(quota correction for tool change): with this function 5 different origins (1,2,3,4,5 are
programmable and passes from an origin to the other wit key
RESET
/
ENTER
. Activating the function of origin
change on the display appear two small arrows and the indication of the selected origin happens with the
switching on of the origin indicators (see paragraph meaning of symbols ). After setting 4 in the first digit on
the right of
//
, will appear
.
and pressing 2 times
RESET
/
ENTER
has to be set the value to be
read in this position of sensor for origin 1, confirm with
RESET
/
ENTER
. It will appear
.
which is the value
to be read for origin 2 in the present position of sensor: set the correct value and confirm with
RESET
/
ENTER
.
It will appears
.
which is the value to be read for origin 3 in the present position of sensor: set the
correct value and confirm with
RESET
/
ENTER
.
For
.
and
.
to proceed as mentioned above.
It means
.
,
.
,
.
,
.
,
.
are references for the calibration, in a certain position of the
sensor, in 5 different origins.
Function
ABS
/
REL
:
Enables the pass from absolute to relative value, pressing key
zero-set temporary the
value to allow a relative shifting. On the display switches on indicator REL to indicate that the current quote is
relative to the zero-point just created. Pressing still key
reappears the absolute value and on the display
switches on the indicator ABS.
Conversion mm/inch:
Pressing key converts the measure from millimetre to inches and back with
indication of inch/mm on the display and a decimal number more then for millimetres. By choosing 5
decimals for millimetres the conversion in inches is not allowed.
Visualization in degrees:
Pressing on key on the display appears Deg to indicate the visualization of
measure in degrees.