L011-80
Rev.
C0
2020-11-05
Page
17
of
17
Recommended Separation Distances:
The mT Controller is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The
customer or user can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF Communications Equipment and the mT Controller as recommended below, according to the maximum output
power of the communications equipment.
Max Output Power
(Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/V1)(Sqrt P)
Separation (m)
80 to 800MHz
D=(3.5/E1)(Sqrt P)
Separation (m)
800MHz to 2.5GHz
D=(7/E1)(Sqrt P)
0.01
0.116667
0.116667
0.233333
0.1
0.368932
0.368932
0.737865
0
1.166667
1.166667
2.333333
10
3.689324
3.689324
7.378648
100
11.66667
11.66667
23.33333