0-20mA (stepped) current output: 3 wire and 4 wire configurations (sink and source).
The 0-20 mA output at different detectors status are define at the below table.
The max load of 0-20 is max 500 ohm at 18-32 VDC. The 0-20 mA is sink isolated and can be
configured as source.
Detector status
Output
Fault
0
mA or 1
mA ±10%
BIT fault
2
mA ±10%
Normal
4
mA ±10%
Alarm
20
mA ±5%
•
Tri-colour LED indication
•
Modbus RTU compatible protocol on RS-485
•
Digital Video: IP network IEEE 802.3 10Base-t
6.5
Mechanical Specifications
•
Size:
7.87 x 5.12× 5.12″ (200×130×130mm)
•
Weight:
o
Detector (Stainless Steel 316): 9.8 lbs. (4.4 kg)
o
Tilt mount (Stainless steel 316): 5.4 lbs. (2.4 kg)
6.6
Environmental Specifications
•
Temperature Range:
o
Operating
-67°F to + 167°F (-55°C to +75°C)
o
Option
-67°F to + 185°F (-55°C to +85°C)
o
Storage
-67°F to + 185°F (-55°C to +85°C)
•
Humidity: up to 99%, non-condensing
•
Ingress Protection: IP66 & 68; NEMA 4X & 6P