![fetatrack 360 User Manual Download Page 18](http://html.mh-extra.com/html/fetatrack/360/360_user-manual_2285142018.webp)
FetalSAT PULSE OXIMETRY (360is)
The FetalSAT system includes the Patient Module, and
the DualSens™ sensor, all of which combine to monitor intrapartum fetal oxygen
saturation (FSpO2), when connected to the Fetatrack 360is.
Intended Use
The FetalSAT Fetal Pulse Oximetry continuously monitors intrapartum fetal oxygen
saturation (FSpO2). It should only be used after amniotic membranes have ruptured
and on a singleton fetus in vertex presentation with a gestational age of at
least 36 weeks. The cervix must be at least 2 cm dilated.
f
etatrack 360is issue 1
17
The FetalSat patient module is connected to
the FetalSat input cable and then to the
Fetatrack 360is before being strapped to the
patients leg.
CAUTION! Do not apply the fetal
spiral electrode when it is
impossible to identify the fetal presenting part.
Apply the DualSens sensor to the fetal scalp. Refer to the DualSens instruction insert
for more specific instructions.
Release the sensor from the applicator.
(Refer to the DualSens Sensor
instruction insert for more specific
instructions regarding sensor removal.)
Connect the DualSens sensor to the
Patient Module.
Fetal oxygen saturation is displayed on
the Fetatrack 360is graphical display
and printed continuously on the UA
scale of the strip chart.