7 - TrOUblesHOOTing
Problem
Cause
resolution
One or both ends of the table will not raise
to the maximum height.
Air has been forced into the hydraulic
reservoir.
Using two people:
●
First Person: Hold down the center and
right pedals (as when lowering)
●
second Person: Repeatedly press the
left pedal (as when raising)
●
First Person: Allow the table to be raised
intermittently by releasing and pressing
the center and right pedals again
●
Continue until the table is at its full
height.
●
Repeat the process if needed.
The hydraulics stick.
The left pedal may have been pumped after
the table top was raised to its maximum
height.
Apply a small amount of downward hand
pressure on the table top to free up the
hydraulics.
The table exhibits any other type of
hydraulic problem.
Contact Ferno Customer Relations (page 2)
or your Ferno distributor.
8 - ParTs anD serviCe
8.1 U.s.a. and Canada
In the United States and Canada, to order parts or for professional
table repair, contact EMSAR
®
- the only agent authorized by Ferno
to manage, service, and repair Ferno products.
Telephone (Toll-Free)
1.800.73.EMSAR
Telephone 1.937.383.1052
Fax +1.937.383.1051
Internet www.emsar.com
Warning
improper parts and service can cause injury. Use
only Ferno parts and Ferno-approved service on
the table.
8.2 Worldwide
To order Ferno parts, and for professional table repair, contact your
Ferno distributor. Your distributor is the only agent authorized by
Ferno to manage, service, and repair Ferno products.
Warning
Modifying the table can cause injury and damage.
Use the table only as designed by Ferno.
15
17
8.3 Parts Diagrams