10
Note: Deleting a position doesn’t create free memory space
for new positions !
When the
del mem
button is pushed longer than 2 seconds, all
stored memory positions are deleted and all memory space comes
available for new measurements.
5.4 Feltest Caliper Profiler PC communication
To capture the measured data from a Caliper Profiler to a
computer, or to chance its settings, you first have to install the
application Feltest Service
(FS)
from included CD-ROM.
A manual and a help file are enclosed in the software.
When the Caliper Profiler is connected to the computer's USB-port
for the first time, the system might require the installation of
drivers from the CD-ROM.
When the Caliper Profiler is connected to a computer, start
FS
and press the "Auto-search" button on the tab "Set-up Profiler".
Available settings:
name
description
range
default
measure-
ments /
second
number of
measurements stored in
memory per second
5 - 25 Hz
10
auto-off
level
when measurements
drop below ….mm/mil
recording stops
0.1-1.5 mm
3.9 – 59 mil
0,15
dimension
unit
measuring in millimeters
or
mils (1/1000”)
mm / mil
mm