F-960 Instruction Manual rev. 6/23/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
33
Phone: (360) 833-8835
[email protected]
www.felixinstruments.com
When Trigger mode is selected, the measurements will display
upon completion in consecutive rows where the user can again
edit and add notes to the data.
Control Panel
The Control Panel Menu allows the user to control the F-960
Ripen It! From a computer using a series of toggle keys seen
below.
By pressing up or down, the user can navigate through the file
menu to change settings on the unit and use the square button
to take a measurement.
Calibration
The Calibration menu enables a two-point calibration process
for the F-960 Ripen It! Including Zero Calibration and Span
Calibration. The set zero process will require known standard
gas of 100% Nitrogen (N2) gas to set a zero baseline. Next the
user will be prompted to connect a known standard gas to set
the span for your sensors. We suggest the following:
C2H4 Set Span: 5 ppm C2H4 standard gas
CO2 Set Span: 95% CO2 standard gas