F-960 Instruction Manual rev. 6/23/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
17
Phone: (360) 833-8835
[email protected]
www.felixinstruments.com
Enter the correct ambient temperature in degrees Celsius,
measured from an external
temperature sensor. To use
the entered temperature for
RH calculations, “use sample
T” should be set to “Yes.”
The current (internal RH) and
corrected (sample RH) are
displayed at the bottom of
the screen.
Setup > GPS
The GPS sensor inside of the F-960 Ripen It! can be used to
record latitude and lon/- 10 meters. The instrument
should be operated outside, without overhead obstruction for
best GPS performance. The Setup > GPS menu turns on/off the
GPS sensor and displays the current GPS data. Data is also
saved to the SD card as a .csv file.
Setup > Language
The Language menu will display options for:
RH Conversion
Enter Sample T(C) 20.0
Use Sample T Yes
(for measurements)
Internal RH (%) 43.6
Sample RH (%) 21.3
GPS
Enable GPS
Yes/No
Acquiring GPS data…..
Longitude 122.558
Latitude 45.59