Calibration
Full set span calibrations should take place annually for the CO
2
sensor and bi-annually for the O
2
sensor. Always set zero before
setting span. If you have not already done so, download the
package setup software from
http://www.felixinstruments.com/support/F-920
on your computer. This package includes the latest firmware
updates, as well as the controller software needed for real time
monitoring and calibration of the F-920. For CO
2
Calibration
standard gases, we suggest 100% CO
2
for the set span
procedure and a verification gas at a lower value, like 20%. For
O
2
Calibration standard gases, we suggest 50% O
2
for the set
span procedure and a verification using ambient air (20.9% O
2
).
1.
Connect the F-920 to computer via a USB cable.
2.
Turn on the F-920.
3.
On the F-920, from the main menu, go to Setup > Mode
and change ‘Connection’ to ‘USB Ctrl’, and ‘Measure’ to
‘Continuous’. Return to the main menu by pressing the
Left arrow.
4.
Open the controller software.
5.
In the top Left of the window, select the F-920 unit
listed.
6.
Click the ‘Calibration’ tab and select the ‘F-920’ sub-tab.