F-920 Instruction Manual rev. 5/10/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
41
Phone: (360) 833-8835
[email protected]
www.felixinstruments.com
FAQ
I have a clogged probe message, what do I do?
-
The F-9xx series will display a clogged probe message
when there is a problem with air flow. This can be
caused potentially by a bad/saturated filter, which is
replaceable. Additionally, check the hosing and probe
for moisture and debris, remove the sampling probe
and see if the clogged probe message goes away.