“Setup > Set Zero” section
of the user manual. Set zero
optionally CAN be done with the Gas Analysis Software (G.A.S.)
utilizing 100% N2 gas (0% CO2 and 0% O2 gas).
After setting zero, select "Span Calibration" and follow the
instructions outlined in the software to complete the Span
calibration process, a diagram outlining how to connect your
standard gas will display.
The timer is defaulted to 3 minutes, which is typically a
sufficient stabilization period, if a longer stabilization period is
needed, increase the timer to 5 minutes. Press start and wait
for the calibration process to complete.