background image

8 / 10

                        O W N E R `S   M A N U A L   F O R   I N C U B A T O R S

 

The   care   you   take   in   reading   and   following   this instruction will   probably determinate the 
satisfactory service you will  receive  from  your Incubator.

UNPACKING

Carefully remove the Incubator  from the shipping case. Preserve all paper work for future reference. 
If damage has occurred from shipment a claim must be filled with the carrier immediately; preserve 
the shipping container for inspection by the carrier. Contact your dealer or Felisa.

INTRODUCTION

FELISA   digital  Incubators    are    equipped   with   a   digital   microprocessor-based control  with 
high quality brilliancy display and with a resolution  of  1 °C.   You can operate your digital Incubator 
with or without timer. Analogic Incubators are equipped with a control knob for the temperature 
control.  All models operates with a  electronic  sensor , inner glass door and  light inside the chamber.

INSTALLATION

Install your Incubator using a connection of the adequate voltage. See the label specification  before 
connecting. The Incubator can   be   installed on any surface or table sufficiently firm and   strong,   
leaving  a  minimum space of 10 cm between the Incubator and any other vertical surface. Keep the 
area around free of any material to allow ventilation in the area of the bottom. For  correct  operation  
it  is  necessary  that  the  user  becomes familiarized with all the available controls and specifications 
shown in each model.

LOAD IN THE CHAMBER

OPERATION TEMPERATURE

Your Incubator is designed to operate satisfactorily to temperatures that do not exceed 100 °C. If this 
temperature  is  not  exceed  the  Incubator    will  provide  a  long  trustable  service.  The  minimum 
operation temperature in the Incubator can’t be less than 5 ºC above  the ambient  temperature. 

 ANALOGIC INCUBATOR  OPERATION

For its operation it is necessary to have on hand a thermometer with a minimum range of 100 °C, 
placed in the thermometer holder (28). Turn on your Incubator with the pilot switch (31). Adjust the 
control knob(33) to the desired temperature. The led indicator (32) will turn on and remain on  until 
reaching the programmed temperature, afterwards it will turn on when the control applies energy to 
the heating element.

You have to wait 35 minutes approximate  in order that the Incubator become stable. Verify if the 
temperature in the thermometer is that you want, if not, readjust the control knob. The experience 
in operation using the control will allow the operator to identify the exact position for different 
temperatures. If you are going to work with the same temperature we suggest that once you have 
the control position for that temperature turn off the Incubator without moving the control knob 
and you will always have the same temperature. 

Summary of Contents for 131AD

Page 1: ...odelos disponibles tambi n con conveccci n mec nica o natural Su construcci n exterior es robusta con gabinete de una sola pieza su panel de control est protegido contra golpes y derrames accidentales...

Page 2: ...UBADORA 0 25 50 75 100 125 150 175 200 225 250 CICLO DIGITAL ANALOGO INCUBADORA 1 3 6 11 12 13 15 16 19 24 2 25 26 27 28 34 35 36 37 39 41 42 43 45 57 58 59 INCUBADORA felisa 32 31 18 47 33 47 52 32 3...

Page 3: ...71 3236 2 71 3336 2 71 3436 37 Cable Conector Bajante Cable 2 71 3177 2 71 3237 2 71 3237 6 71 3437 38 Cable Conector Bajante Connector 1 71 3238 1 71 3238 39 Cable Conector Connector 1 71 3239 1 71...

Page 4: ...mente encender cada vez que el control aplique energ a al elemento calefactor para mantener la temperatura fijada Espere un tiempo de aproximadamente 45 minutos a que la Incubadora se estabilize para...

Page 5: ...ej desactivadoeltimerautom ticamentepasar alaconfiguraci ndelaAlarmaauditiva Programaci n del timer el display presentar el mensaje hh y enseguida las horas que tiene programado el sistema para ajusta...

Page 6: ...as mediciones del display y de suterm metro siexistealgunadiferenciasignificativa deber procederdelasiguientemanera Apague la Incubadora con el interruptor 31 Con la tecla de men oprimida encienda la...

Page 7: ...zar cualquier labor de mantenimiento desconecte el cable de alimentaci n Para preservar en buen estado los elementos calefactores no derrame ninguna soluci n dentro de la c mara o introduzca objetos m...

Page 8: ...n the area of the bottom For correct operation it is necessary that the user becomes familiarized with all the available controls and specifications shownineachmodel LOADINTHECHAMBER OPERATIONTEMPERAT...

Page 9: ...SP Set Point for two seconds and then theconfiguretemperature Withtheincreaseordecreasebuttonsyoucanmodifythisparametertothe desired operation temperature then press the menu button to continue with...

Page 10: ...rmometer temperature against the display if there is a difference you have to calibrate your Incubator Turn off the Incubator with the on off switch 31 with the menu button pressed turn on the Incubat...

Page 11: ...olucrado Felisa se reserva el derecho de hace cambios o modificaciones en sus productos sin previo aviso con el findemejorarsupresentaciony ooperacion All products manufactured by Felisa are guarantee...

Reviews: