IRB20 Inverted Microscope Bulb & Fuse Replacement

12

Fein 

TROUBLE

POSSIBLE CAUSE

SOLUTION

Coarse focusing knob 

is too tight.

Tension knob is too loose.

Tighten it a little.

Stage declines itself, 

cannot stay on the 

focal plane when 

observing.

Tension knob is too loose.

Tighten it a little.

Coarse focusing knob 

can’t rise.

The limit stop knob is locked.

Loosen the knob.

Coarse focusing knob 

can’t decline.

The base of the condenser is too low.

Raise the base.

Cannot move the 

slide smoothly.

The slide is not fixed correctly.

Adjust it 

The movable specimen holder is not fixed 

properly.

Adjust it

The image moves 

obviously when 

touching the stage.

The stage is fastened incorrectly.

Fasten the stage correctly.

The LED light does 

not work.

No power supply.

Check the connection of the 

power cable.

The LED bulb is not installed correctly.

Install it correctly.

The LED bulb is burnt out.

Replace it.

The bulb burns out 

frequently.

The wrong bulb is being used.

Replace it with the correct bulb.

The illumination is 

not bright enough.

The wrong bulb is being used. 

Replace it with the correct bulb.

The illumination adjustment is not set in a 

bright position.

Adjust the illumination knob to 

intensify the light.

Troubleshooting

Summary of Contents for RB20-HD

Page 1: ...Fein RB20 Biological Microscope User s Manual info feinoptic com...

Page 2: ...Fein Microscope Components Eyepieces Condenser Objective Lens 1 Coarse Adjustment Fine Adjustment Tension Adjustment Microscope Frame Carrying Handle Condenser Adjustment Knob...

Page 3: ...epiece Mechanical Stage Fein Microscope Components 2 Lock Screw Fixed Koehler Illumination Focus Arm Trinocular Head Vertical Adjustment Knob Horizontal Adjustment Knob Power Switch Light Adjustment K...

Page 4: ...ck 1 and the frame at the front 2 as shown in the image at right Set the power switch to off O before replacing a bulb or fuse and wait until the lamp is cool The power switch is located on the side o...

Page 5: ...4 Fein Microscope Assembly Follow the numbers below when assembling the microscope Eyepiece Condenser Objectives Power Cord 1 2 3 4...

Page 6: ...il the condenser is aligned with the base and rotate the condenser to make the handle face forward Tighten the condenser lock screw 2 of condenser then raise the condenser with the up down knob to the...

Page 7: ...ntil the eyepiece is flush with the eyepiece tube Tighten the eyepiece with M2 5 inner hexagon lock screw 3 1 3 2 Assemble the Eyepiece Take off the eyepiece tube cover 1 Insert the eyepiece 2 into th...

Page 8: ...wly move the specimen holder lever back into place This will secure the specimen slide Rotate the X and Y axis knobs 3 on the stage until the center of the specimen slide is aligned with the objective...

Page 9: ...e eyepieces 2 to fit your personal interpupillary distance The scale between the eyepieces 1 notes the interpupillary distance setting It s a good idea to note this number when working in a setting wh...

Page 10: ...r use wipe the lens with a tissue moistened with microscope cleaning solution Using Filters Filters can increase the contrast in microscopy images The filter sits on top of the light as shown at the r...

Page 11: ...e microscope is not in use the power cord can be placed in the storage box Press the position 1 of the storage box cover with thumb When you hear a click push up and put the power cord into the storag...

Page 12: ...over glass The cover glass is not standard Use a standard cover glass with thickness of 0 17mm The slide is upside down Adjust the slide The immersion oil has accumulated on the dry objective Clean th...

Page 13: ...xed correctly Adjust it The movable specimen holder is not fixed properly Adjust it The image moves obviously when touching the stage The stage is fastened incorrectly Fasten the stage correctly The L...

Page 14: ...rrying handle side is down Locate the 4 rubber feet on the underside of the microscope Remove the rubber feet exposing the 4 screws securing the bottom plate of the microscope Gently lay microscope on...

Page 15: ...th all 4 screws removed carefully remove the bottom plate exposing the internal components of the microscope Locate the 2 screws securing the LED bulb in place Using a screw driver remove the 2 screws...

Page 16: ...the LED bulb replaced position the bottom of the microscope plate back onto the microscope frame Using a screw driver secure the bottom plate with the 4 set screws Place the 4 rubber feet over the to...

Reviews: