12
F-049 GB Rev00 02.05.2019
get up® Instruction Manual
9.3.2 Raising the Handle
1.
Grab the handle with one hand.
2.
Lightly pull down the triangle handle to
tighten the strap.
3.
Push and hold the belt buckle with the other
hand.
The belt buckle is now unfastened.
4.
Reduce the fraction on the triangle handle
until the handle is lifted to the desired
position.
5.
Release the button on the belt buckle.
The belt buckle snaps audibly.
6.
Make sure that the handle has been raised
as desired.
Abb.5: Raising the Triangle Handle
10 Preparation
The get up® must be disinfected for cleaning.
The following warnings must be observed:
ATTENTION
Loss of Efficacy Due to Inappropriate Aids
Poorly absorbent aids lead to insufficient
disinfection.
Only use absorbent material,
e.g. fleece.
ATTENTION
Environmental Damage Due to Disinfectants
If surface disinfectants find their way into soils
and waters, the environment can be put at risk.
Always dispose of surface disinfectants
without polluting surface waters or the
sewer system.
Follow the disposal instructions of the
disinfectant manufacturer.
WARNING
Chemical Burn on Hair or Skin
If surface disinfectant finds its way to unpro-
tected hair or skin, chemical burns can occur.
Always wear protective gloves when in
contact with surface disinfectant; Protective
gloves made of nitrile or butyl rubber are
particularly suitable.
If brought into contact with hair or skin:
Immediately wash the surface disinfectant
off the skin and hair using tap water of
drinking quality.
WARNING
Serious Eye Damage
1
2
3
4
5
Summary of Contents for get up
Page 1: ...get up The Swiveling Hold System for Radiology Instruction Manual F 049 GB Rev00 02 05 2019 ...
Page 21: ...get up Instruction Manual F 049 GB Rev00 02 05 2019 21 18 Notes ...
Page 22: ...22 F 049 GB Rev00 02 05 2019 get up Instruction Manual ...
Page 23: ...get up Instruction Manual F 049 GB Rev00 02 05 2019 23 ...