![FE ZAJ Service Manual Download Page 20](http://html1.mh-extra.com/html/fe/zaj/zaj_service-manual_4600110020.webp)
14
TN512955-E
No.
Name
Man. Drawing No.
Remarks
Recommended
replacement cycle
1
Restrictor
TK708941C1
Reference gas joints
2 years
2
Filter
TK708922P2
Reference gas joints
2 years
3
Packing
TK722676P4
Near the filter
2 years
4 Packing
TK722676P4
For
restrictor
2
years
5 Fuse
757642
2A,
6, 4 x 30 mm
2 years
* Recommended replacement cycle for maintenance parts is at least 2 years. Prepare maintenance
parts every 2 years.
* Maintenance parts including No. 1 to No. 4 used near the reference gas should be replaced every 2
years. However, if there is no problem with reference gas flow (a constant flow of 5 to 20 mL is
maintained) or indication (does not fluctuate), there is no need to replace them.