3-2. Output
adjustment
(4 to 20mA, 0 to 1V)
To perform output adjustment, connect a digital voltmeter to the output terminal and set the
measurement screen in the Factory Adjustment mode.
Caution:
Perform the output adjustment within the range of 4 to 2mA and 0 to 1V on the
Adjustment screen. If other adjustment screen appears and you are at a loss what to do
on the screen, press the ESC key several times to change it to the Measurement mode.
In this case, don’t press the ENT key. Otherwise the adjustment contents would change,
and disable adjustment.
• Factory
adjustment
mode
Press the MODE key several times on the Measurement screen, and the User Adjustment screen
appears. Press the ZERO key 5 times on the User Adjustment screen, and the Factory Adjustment
mode will be selected. This mode contains 4 to 20mA output, 0 to 1V output, coil pulse, manual gain,
synchronous pulse, zero memory, span memory, magnet, reset, and peak code input. Be careful not
to get in items other than “4 to 20mA output” and “0 to 1V output”.
In 4 to 20mA output and 0 to 1V
output, be sure to press the ENT key to save the change that is adjusted by (
▲
) or (
▼
) key. After
adjustments, press the ESC key to escape to the Measurement mode.
0-1V OUT 0V
Press
▲▼
key
MEAS MODE
H:ON R:OFF L:OFF A:OFF
ADJUST MODE
>
MOVE AVE
▲
CLOCK
Factory Mode 0-1V
▲
4-20mA
▼
Sync period
0-1V OUT 0V
▼
1V
ENT
ESC
MODE
ESC
Press the
key several times to get in the Adjustment mode.
ZERO
ESC
Press the key 5 times.
Move “4-20mA” or “0-1V” up one line by (
▲
) or (
▼
) key. Press the
( > ) key to enter the Output Adjustment screen.
Select either 0V (zero) or 1V
(span) and press the ( > ) key.
Adjust the output by (
▲
) or (
▼
)
key while reading the digital
voltmeter and press the ENT key.
ESC
* Adjustment of 4 to 20mA should be performed in the same procedure as in 0 to 1V.
Summary of Contents for ZAJ
Page 1: ...Service Manual TN512955b E PARAMAGNETIC OXYGEN ANALYZER TYPE ZAJ...
Page 24: ...18 TN512955 E Appendix 3 1 CPU printed board J6 J5...
Page 25: ...TN512955 E 19 Appendix 3 2 A D printed circuit board J11 J15 J13 J12 J14...
Page 26: ...20 TN512955 E Appendix 3 3 Display output printed circuit board VR1...
Page 28: ......