FASTLITE MOZZA Operating Manual Download Page 8

Mozza manual

Chapter 2 : Mozza Theory

diffracted beam is eventually taken into account when the reconstruction of the spectrum is
done.

FASTLITE - Mozza manual V1.2

7/

21

Summary of Contents for MOZZA

Page 1: ...MOZZA SYSTEM OPERATING MANUAL V1 2 V1 2 FASTLITE info fastlite com...

Page 2: ...Mozza manual FASTLITE Mozza manual V1 2 1 21...

Page 3: ...starting 9 4 1 Mozza application software 9 4 2 Connections and starting 9 5 Software guide 11 5 1 Menu area 12 5 1 1 File menu 12 5 1 2 Device menu 12 5 1 3 Operations menu 13 5 1 4 View menu 14 5 1...

Page 4: ...d on an acousto optic tunable filter AOTF and is specifically well adapted to pulsed sources such as femtosecond amplifiers OPO or OPA with repetition rate up to 640 kHz however it can also be used wi...

Page 5: ...igent operator understanding at least the basic principles and limitations of the technique We hope the reading of this document will give you the necessary insights and allow you to take full advanta...

Page 6: ...provides many advantages over other scanning tech niques Using a solid state crystal and selecting the wavelength electronically no moving part is used in this setup as a consequence it is reliable f...

Page 7: ...most of the transmitted beam is removed from the detector a broadband polarizer is used to select P polarization which corresponds to the polariza tion of the diffracted beam the transmitted beam bei...

Page 8: ...Mozza manual Chapter 2 Mozza Theory diffracted beam is eventually taken into account when the reconstruction of the spectrum is done FASTLITE Mozza manual V1 2 7 21...

Page 9: ...zza crate is re usable and should be kept for further servicing 3 2 Mozza Serial Number The Serial Number provides essential information about the MOZZA and is indicated on the MOZZA test sheet and on...

Page 10: ...ed at FLASHDRIVE setup mozza 360 r241 exe Windows 7 and 8 are supported The software is installed by executing the file leave all default options This will create the icon 4 2 Connections and starting...

Page 11: ...s are cooled down with Peltier modules the red light indicates that they have not reached correct temperatures after 30 seconds 1 minutes the red light should not be blinking indicating the correct be...

Page 12: ...nfiguration of the software Figure 5 1 Mozza software main window Menu area gives access to File Device Operations View and Help settings as de scribed in section 5 1 Basic Settings allows to select t...

Page 13: ...rent point in the scan Graph is a selectable display area The available contents depend on the value of the Graph Selection control A description of these displays is included in the following 5 1 Men...

Page 14: ...the computer Mozza0 xxxx is the list of all Mozza hardware detected by the software Figure 5 4 Monitoring window 5 1 3 Operations menu Figure 5 5 Operations menu Alignment allows the user to remove a...

Page 15: ...nt of the Trigger delay parameter must be done so that the background can be processed properly by the software green points on the figure before the pulse and the sample points red points on the figu...

Page 16: ...is connected or disconnected it gives the trigger frequency the approximate scan duration and the real duration of the last scan Figure 5 10 Status bar 5 3 Basic Settings start is the starting point...

Page 17: ...icate Waiting instead of Start Abort will finish the measurement immediately 5 4 Advanced Settings Figure 5 12 Advanced settings Trigger source allow the user to choose between Internal with continuou...

Page 18: ...indicator of the lock in mode 5 5 Panel The panels show the recorded data Displayed data depends on the mode single frequency or scan and on the option chosen in graph selection The left panel display...

Page 19: ...e MOZZA requires a linearly vertically polarized beam and a maximum of 10 J focused on the slit the user can put more power if working with collimated beam 6 1 Basic alignment When power of the infrar...

Page 20: ...lacing 2 iris Then place the focusing optic lens or concave mirror and ensure that the visible light is going in the right direction For this purpose the cover can be opened the user should provide th...

Page 21: ...arrives at least 40 s before the pulse If the laser repetition rate is high frep 1kHz the trigger laser must arrive at the same time as the laser pulse When working with external trigger choose Trigge...

Page 22: ...ctrometer and wait few seconds Verify that the green and red LEDs on the spectrometer back slide are lit If not check that it is correctly listed as Mozza Spectrum Analyzer in the device manager 7 2 N...