FN ICON CALIBRATION AND REBUILD INSTRUCTIONS 

 

 

    Roseville, MN 

     Ph: 1-800-444-2373 
     Fax: 651-645-7390 

     www.fastestinc.com 

                                                             Page 

1

 of 

5

                                                                                                                                                                                                    WP158 Rev C, 12/13/2021  

 

 

        

WARNING

 

Use only factory authorized replacement seals and parts. Use 
of unauthorized parts/seals can cause failure resulting in 
damage or injury.

 

THREADED  

GRIP COLLETS 

SQUEEZE HANDLE 

(RELEASED 

POSITION COLLETS 

ENGAGED) 

PNEUMATIC 

CONNECTOR 

MEDIA PORT 

MAIN SEAL 

O-RING 

SQUEEZE 

HANDLE (INSERT 

AND REMOVE 

POSITION) 

GENERAL INFORMATION:

 

The 

FasMate

 

is designed to provide a safe, reliable leak-tight seal and 

connection when properly maintained and used. A flat, smooth, sealing 
surface perpendicular to the connector axis is required around the female 
test port. Verify the connector seal mates well with the sealing surface.  
Improper mating with a test piece can result in poor sealing, leakage and risk 
of connector failure resulting in damage or injury. 

MAINTENANCE:

 

User must establish a regular interval for maintenance as determined by the 
user media and operational environment. 
 
Periodically inspect the 

FasMate

 

and its seal for wear, damage and proper 

operation. Inspect also for lubrication needs or corrosion. Repair or replace as 
required to assure proper sealing, function, and safety. 
  
Difficulty  of  operation  indicates  a  need  for  inspection,  lubrication,  repair  or 
other maintenance. 

 

 

Parts  and  instructions  available  through 

FasTest

  or  your 

FasTest

 

distributor. 

NOTE: Any field modification of 

FasTest

 connectors voids Factory war-

ranty. 

FasTest, Inc. Product Warranty

 

FasTest, Inc. warrants its products against defects of workmanship and/or material for 12 months from the date of the 
sale by FasTest, Inc.  This warranty is void if the product is misused, tampered with or used in a manner that is not in 
accordance with FasTest, Inc. recommendations and/or instructions. FasTest, Inc. is not liable for consequential or 
other damages including, but not limited to, loss, damage, personal injury, or any other expense directly or indirectly 
arising from the use of or inability to use its products either separately or in combination with other products. ALL 
OTHER  WARRANTIES  EXPRESSED  OR  IMPLIED,  WHETHER  ORAL  OR  WRITTEN,  INCLUDING  BUT  NOT 
LIMITED TO  
WARRANTIES  OR  MERCHANTABILITY  OR  FITNESS  FOR  A  PARTICULAR  PURPOSE  ARE  EXPRESSLY 
EXCLUDED. 
Remedy under this warranty is limited to replacement of the product or an account credit in the amount of the original 
selling price, at the option on FasTest, Inc.  All allegedly defective products must be returned prepaid transportation to 
FasTest, Inc. along with information describing the products performance, unless disposition in the field is authorized 
in writing by FasTest, Inc. 

FasMate FN Series w/ICON 

Description: Lever & Pneumatic operated Intelligent 
Connectors for Female Threads. 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 

 
FasMate

 Connector provides a reliable leak-tight connection that grips and 

seals into female threads. 
Please thoroughly read and understand these operating instructions prior to 
operating  connector.  The  use  of  pressurized  media  for  sealing  and  testing 
requires a thorough understanding of the 

FasMate

 

Operating Instructions. 

 

 

Operation 

 

General Information 

 

Maintenance 

 

Warranty 

Other FN Resources: 

WP101 Operating Instructions 
WP118 Rebuild Instructions 
WP119 Main Seal Replacement Instructions 

Summary of Contents for FN Series

Page 1: ...nts its products against defects of workmanship and or material for 12 months from the date of the sale by FasTest Inc This warranty is void if the product is misused tampered with or used in a manner that is not in accordance with FasTest Inc recommendations and or instructions FasTest Inc is not liable for consequential or other damages including but not limited to loss damage personal injury or...

Page 2: ...STEPS 3 6 BEFORE STARTING CALIBRATION Note Point A1 is a preset distance from the rest position A As the shaft is extended from rest it will begin flashing yellow within the good zone A good connection is only indicated when the shaft retracts from beyond the upper limit B If a good connection is made and it s just to the right of Point A1 it could travel through point A1 once test pressure is int...

Page 3: ...t a good connection on an analog Sure SealTM FasMate the voltage must be above the short connect threshold voltage VSC and below the cross thread detection voltage VCT This calibration procedure is a guideline for determining these two voltages but exact values depend on the user application Lever Control FNL with Analog 1 Rotate red housing to desired location and lock in place with set screw a R...

Page 4: ...st voltage observed for a good connection minus some safety margin 7 A good connection is indicated by the output voltage settling in the range between Vsc and VCT 8 Once the range is determined a 24VDC signal may be applied to pins 2 and 5 to provide red yellow green LED indication of connection quality When using Analog please note that the shielded cable improves noise single termination 0 10VD...

Page 5: ...w pressure to remove the pressure piston The spring and actuator are located behind that 4 Remove o ring from pressure piston and replace 5 Install new spring and actuator 6 Coat outside of o ring with petroleum jelly 7 Push pressure piston back inside red housing Press in as squarely as possible 8 Apply Loctite 242 on screws and secure new PCA module tighten to 5 in lbs 8 Re install red housing o...

Reviews: