Fasep 2000 srl
Rev. 1.2
Balatron 331.G3: User’s Manual
May 31, 2013
4-2
F
IG
.
11
F
IG
.
12
4.2
How to check the calibration of Wheel Balancer (USER)
ATTENTION
select mode
CAL USR
Press <+/-> to select
CAL TEST
Place a wheel on the flange ( Fig. 11 ).
Put the calibration weight ( Fig. 12 ).
Press
<FINE>
to see actual values. 160-
0 (±3) is correct result.
When all LED (left side) are lit,
calibration weight must be at exactly 6
o’clock.
If no, press
<FUNC>
to calibrate
position.
Turn the wheel until the calibration
weight is located at 6 o’clock.
Press
<MODE/ESC>
to return to normal
balancing mode.
NOTE
technical calibration is only available for qualified personnel.