fasciotens Carrier HC010 Instructions For Use Manual Download Page 8

8

Instructions for Use

Carrier

Assembly 

 

Carrier

 

HC010

4.   Open the eccentric handle by flipping the clamping 

lever. 

The word "open" will now be visible. Make sure 

that the locking device on the eccentric handle engages 
correctly.

5.   Move the eccentric handle on the post up or down 

to 

position the product based on how corpulent the patient 
is. We recommend positioning the ball holder 15-20 cm 
above the abdomen.

6.   Remove the crossbar from the packaging 

and

 

insert 

the crossbar into the eccentric handle.

7.   Move the crossbar in the desired direction

 to find the optimal setting depending on the location of the defect. 

8.   Secure the post in the eccentric handle 

by flipping the clamping lever. The word "closed" will now be visible.

The post and the crossbar can cause pressure damage. Make sure that there is sufficient space between the 
patient and the crossbar.

Summary of Contents for Carrier HC010

Page 1: ...Carrier www fasciotens de Instructions for Use Carrier...

Page 2: ...ontrary to its intended purpose or incorrect handling Any significant incidents occurring in connection with the product must be reported immediately to fasciotens GmbH and the responsible national au...

Page 3: ...structure Carrier HC010 6 Assembly Carrier HC010 7 Product structure Carrier HC020 10 Product assembly Carrier HC020 11 Combination with Hernia 14 Processing instructions Carrier 15 Service life 15 Pr...

Page 4: ...perating the product or have not receive proper training in its use In addition liability is transferred to the user in case of improper use or use that is not within its defined intended purpose Prio...

Page 5: ...uman medical purposes and is used during surgery The product is approved for use in combination with Hernia Indications Combination with products Combination with operating tables or standard rails Co...

Page 6: ...ust be sterilised in the hospital before each use in the operating room Please follow the processing instructions Please follow the storage instructions for the product Carrier must be sterilised befo...

Page 7: ...e them on the instrument table Make sure that the clamp opening at the lower end of the post is fully open 2 Placethepostonthestandardrailoftheoperatingtable 3 Lock the post on the standard rail of th...

Page 8: ...e patient is We recommend positioning the ball holder 15 20 cm above the abdomen 6 Remove the crossbar from the packaging and insert the crossbar into the eccentric handle 7 Move the crossbar in the d...

Page 9: ...crossbar has an individually adjustable angle which can be set by means of a rotary wheel in front of the ball holder Before each use check that the product is intact and make sure that the post and...

Page 10: ...cturer and must be sterilised in the hospital before each use in the operating room Please follow the processing instructions Please follow the storage instructions for the product Carrier must be ste...

Page 11: ...ng at the lower end of the post is fully open 2 Turn the handwheel into the hole provided for this purpose at the lower end of the central post 3 Place the post on the standard rail of the operating t...

Page 12: ...ts of the post by turning the handwheel 7 Guide the eccentric handle P4 onto the top of the post extension and move it to the desired position 8 Take the crossbar and insert it into the open eccentric...

Page 13: ...entric handle by flipping the clamping lever The word closed will now be visible on the eccentric lock Before each use check that the product is intact and make sure that the post and the crossbar are...

Page 14: ...button on the clamping lever and opening the clamping lever at the same time Hernia can now be inserted into the ball holder from above Fasten the ball adapter by inserting it into the retaining brack...

Page 15: ...on and the processing of contaminated instruments Preparation We recommend processing the contaminated instruments as soon as possible after they have been used They should be transported in a sealed...

Page 16: ...r s instructions In recent studies on de contamination procedures against infectious prion proteins the most effective methods have been successive treatment with an alkaline detergent pH 11 and disin...

Page 17: ...to fasciotens GmbH for repair must also be sterilised beforehand individual packaging see section Sterilisation Please use the template for returns from your AEMP CSSD folder for this purpose After ea...

Page 18: ...age instructions In principle Carrier should be stored in a clean cool and dry location be protected from mechanical damage not be dropped and be handled with care The generally applicable regulations...

Page 19: ...onfirmation must be enclosed with the return shipment Carrier We hereby confirm that the enclosed instrument set has been correctly disinfected cleaned and sterilised Instrument set Confirmation stick...

Page 20: ...rranty period of 24 months applies to our products Should any initial defect occur in your product within this period please inform our Customer Support immediately If there are any deficiencies that...

Reviews: