fasciotens Carrier HC010 Instructions For Use Manual Download Page 17

17

Instructions for Use

Carrier

Processing

 

Carrier

 

3. Machine cleaning

Place the instruments on sieve baskets or racks suitable for cleaning purposes and ensure that the water jet can reach all 
areas. A 0.5% alkaline cleaner (e.g. Deconex 28 Alka One, Borer Chemie) should be used according to the manufacturer's in-
structions for use. Pay attention to the correct dosage! 

Carrier

 is approved for alkaline cleaning. Acidic cleaning 

agents and disinfectants must not be used.

The device manufacturer's instructions must also be followed. A typical cycle should include the following steps:

• 

At least 2 minutes of pre-washing with cold water

• 

Treat for at least 5 minutes with 0.5% alkaline cleaner (e.g. Deconex 28 Alka One, Borer Chemie) 
below protein coagulation temperature (<60 °C)

• 

At least 3 minutes neutralisation with warm, fully demineralised water (>40 °C)

• 

At least 2 minutes of rinsing with warm, fully demineralised water (>40 °C) to remove cleaning agent

• 

Hot wash disinfection (max. 93 °C)

• 

Dry cycle (max. 60 °C)

The precise parameters may vary depending on the load and the program.

The instruments should be removed from the machine immediately after the end of the program and cooled to room tem-
perature. They should not be left in the washing machine after the washing process.

After cleaning, check all parts for visible dirt (especially in cannulations and blind holes). If necessary, repeat the cycle or 
clean manually. All articulated parts must be dried after cleaning. Inspect parts for damage that could impair their function-
ality. Damaged and defective instruments must be identified and replaced. Instruments to be returned to fasciotens GmbH 
for repair must also be sterilised beforehand (individual packaging, see section Sterilisation). Please use the template for 
returns from your AEMP/ CSSD folder for this purpose. 

After each cleaning and cooling cycle, areas such as joints, sutures, etc. must be treated with suitable care products (medic-
inal white oil).

Sterilisation

Instruments can be sterilised individually wrapped (in a standard sterilisation pouch) in appropriate container systems or 
general purpose sterilisation containers. The containers should not be overloaded. Please pay attention to the manufactur-
er's instructions!

Sterilisation should be carried out in accordance with a validated procedure using steam with fractionated pre-vacuum (e.g. 
steriliser according to EN 285 and validated according to DIN EN ISO 17665-1). An exposure time of at least 5 minutes must be 
maintained at a temperature of 134 °C. All joints and eccentric closures must be open during sterilisation.

Final instructions

These instructions have been determined by fasciotens GmbH to be suitable in order to prepare 

Carrier

  for re-

use. The processor is responsible for ensuring that the processing carried out with the equipment, materials and personnel 
used in the processing facility achieves the desired results. This normally requires validation and routine monitoring of the 
process. Likewise, any deviation from the instructions provided should be carefully evaluated by the processor to make sure 
they are effective and to identify any potential adverse consequences.

We hereby confirm that no products leave our facility until appropriate quality control has been performed. Nevertheless, 
issues are possible. Please check the goods are complete and functional and inform us immediately if you have any com-
plaints. Please do not use rejected goods!

Summary of Contents for Carrier HC010

Page 1: ...Carrier www fasciotens de Instructions for Use Carrier...

Page 2: ...ontrary to its intended purpose or incorrect handling Any significant incidents occurring in connection with the product must be reported immediately to fasciotens GmbH and the responsible national au...

Page 3: ...structure Carrier HC010 6 Assembly Carrier HC010 7 Product structure Carrier HC020 10 Product assembly Carrier HC020 11 Combination with Hernia 14 Processing instructions Carrier 15 Service life 15 Pr...

Page 4: ...perating the product or have not receive proper training in its use In addition liability is transferred to the user in case of improper use or use that is not within its defined intended purpose Prio...

Page 5: ...uman medical purposes and is used during surgery The product is approved for use in combination with Hernia Indications Combination with products Combination with operating tables or standard rails Co...

Page 6: ...ust be sterilised in the hospital before each use in the operating room Please follow the processing instructions Please follow the storage instructions for the product Carrier must be sterilised befo...

Page 7: ...e them on the instrument table Make sure that the clamp opening at the lower end of the post is fully open 2 Placethepostonthestandardrailoftheoperatingtable 3 Lock the post on the standard rail of th...

Page 8: ...e patient is We recommend positioning the ball holder 15 20 cm above the abdomen 6 Remove the crossbar from the packaging and insert the crossbar into the eccentric handle 7 Move the crossbar in the d...

Page 9: ...crossbar has an individually adjustable angle which can be set by means of a rotary wheel in front of the ball holder Before each use check that the product is intact and make sure that the post and...

Page 10: ...cturer and must be sterilised in the hospital before each use in the operating room Please follow the processing instructions Please follow the storage instructions for the product Carrier must be ste...

Page 11: ...ng at the lower end of the post is fully open 2 Turn the handwheel into the hole provided for this purpose at the lower end of the central post 3 Place the post on the standard rail of the operating t...

Page 12: ...ts of the post by turning the handwheel 7 Guide the eccentric handle P4 onto the top of the post extension and move it to the desired position 8 Take the crossbar and insert it into the open eccentric...

Page 13: ...entric handle by flipping the clamping lever The word closed will now be visible on the eccentric lock Before each use check that the product is intact and make sure that the post and the crossbar are...

Page 14: ...button on the clamping lever and opening the clamping lever at the same time Hernia can now be inserted into the ball holder from above Fasten the ball adapter by inserting it into the retaining brack...

Page 15: ...on and the processing of contaminated instruments Preparation We recommend processing the contaminated instruments as soon as possible after they have been used They should be transported in a sealed...

Page 16: ...r s instructions In recent studies on de contamination procedures against infectious prion proteins the most effective methods have been successive treatment with an alkaline detergent pH 11 and disin...

Page 17: ...to fasciotens GmbH for repair must also be sterilised beforehand individual packaging see section Sterilisation Please use the template for returns from your AEMP CSSD folder for this purpose After ea...

Page 18: ...age instructions In principle Carrier should be stored in a clean cool and dry location be protected from mechanical damage not be dropped and be handled with care The generally applicable regulations...

Page 19: ...onfirmation must be enclosed with the return shipment Carrier We hereby confirm that the enclosed instrument set has been correctly disinfected cleaned and sterilised Instrument set Confirmation stick...

Page 20: ...rranty period of 24 months applies to our products Should any initial defect occur in your product within this period please inform our Customer Support immediately If there are any deficiencies that...

Reviews: