Model 35 Viscometer Instruction Manual
208878
Revision N, February 2013
15
5.1
Operating the Model 35A and 35SA
The Model 35A and 35SA viscometers operate at six speeds, ranging from 3 rpm
to 600 rpm. To select the desired speed, set the speed switch (located on the right
side of the base) to the high or low speed position as desired. Then turn the motor
on and move the gear shift knob (located on the top of the instrument) to the
position that corresponds to the desired speed.
Table 5-1 lists the positions for the viscometer switch and the gear knob
combinations to obtain the desired speed. The viscometer gear shift knob may be
engaged while the motor is running. Read the dial for shear stress values.
Table 5-1 Six-Speed Testing Combinations for Models 35A and 35SA
Speed RPM
Viscometer Switch
Gear Shift Knob
600
High
Down
300
Low
Down
200
High
Up
100
Low
Up
6
High
Center
3
Low
Center
5.2
Operating the Model 35A/SR-12 and 35SA/SR-12
The Model 35A/SR-12 and 35SA/SR-12 have twelve speeds for testing
capabilities. To achieve this broader testing range from 0.9 rpm to 600 rpm, an
additional gear box shift lever is used; it is located on the right side of the gear box.
See Figure 5-1. Move this lever to the left or right as determined from Table 5-2.
Never change the gear box shift lever while the motor is running.
Changing it while the motor is running will result in gear damage
Only the viscometer gear shift knob (on top of the instrument)
can be changed while the motor is running.